Update of the practical guide for validation studies of automated cuff blood pressure monitors according to the Universal Standard by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018) and Amendments 1:2020 and 2:2024.
George S Stergiou, Ariadni Menti, Roland Asmar, Grzegorz Bilo, Angeliki Ntineri, Paolo Palatini, Gianfranco Parati, Andrew Shennan,
Hypertension Center STRIDE-7, National and Kapodistrian University of Athens, School of Medicine, Third Department of Medicine, Sotiria Hospital, Athens, Greece. · Institute of Cardiovascular Science, University College London and the National Institute for Health Research University College London Hospitals Biomedical Research Centre, UK.
Key points
From the paper's conclusion- In 2018, the Association for the Advancement of Medical Instrumentation (AAMI), the European Society of Hypertension (ESH), and the International Organization for Standardization (ISO) introduced the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) to establish globally accepted requirements for validation studies.
Abstract
Validated blood pressure (BP) monitors are essential for the accurate diagnosis and management of hypertension. In 2018, the Association for the Advancement of Medical Instrumentation (AAMI), the European Society of Hypertension (ESH), and the International Organization for Standardization (ISO) introduced the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) to establish globally accepted requirements for validation studies. This standard is developed specifically for cuff BP devices and is inadequate for cuffless ones, which require different validation procedures. In 2018, the ESH Working Group on Blood Pressure Monitoring published a practical guide and checklist for performing and reporting validation studies of cuff BP devices. Thereafter, two recent amendments (1:2020 and 2:2024) introduced additional requirements for limb and cuff size distribution to ensure that the entire limb range per cuff is adequately assessed. Despite these major initiatives, regular reviews of published validation studies performed by STRIDE BP have identified persistent violations of the validation requirements and incomplete reporting of important data. This statement, prepared by STRIDE BP and the ESH Working Group on Blood Pressure Monitoring, provides an updated guide and checklists, which are complementary to our 2019 guide and aim at assisting investigators and reviewers in performing and reporting high-quality validation studies of cuff BP devices that comply with the latest requirements by the AAMI/ESH/ISO Universal Standard.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical investigation of the iHealth Track KN-550BT upper-arm blood pressure monitor in pregnant patients according to the ISO 81060-2:2018+AMD1:2020/AMD2:2024.Blood pressure monitoring · 2026
- Validation of the ARM-30F-XL arm blood pressure monitor with multiple cuff sizes according to the International Standard ISO 81060-2:2018+Amd 2:2024.Blood pressure monitoring · 2026
- Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.Blood pressure monitoring · 2026
- Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020)Blood Pressure · 2026
- Toward patient-centered blood pressure monitoring: validation of a novel low-pressure and rapid oscillometric method according to the ISO 81060-2:2018/Amd1:2020/Amd2:2024 standard.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- To be (IIb), or not to be (IIb) - That is the current status of cuffless blood pressure monitoring device.Clinical hypertension · 2026