B-278 Markedly Elevated Lactate Dehydrogenase as an Under-Recognized Interference in Enzymatic Chemistry Assays Utilizing NAD/NADH-Coupled Reactions
Annalara Fischer, Cierra Sharp, Nicole Pozzi, Saeed Jortani
Key points
From the paper's conclusion- Given the assay’s NAD/NADH-coupled enzymatic methodology, interference from considerably elevated endogenous LDH was suspected.
Abstract
Background Many automated chemistry assays utilize enzymatic reactions coupled to NAD/NADH conversion with spectrophotometric measurement. While manufacturers validate common interferents, the impact of extreme endogenous enzyme elevations is not always fully characterized. Lactate dehydrogenase (LDH), which catalyzes NAD/NADH interconversion, may interfere with assays of similar design when present at markedly elevated concentrations. We describe two cases in which severe LDH elevation produced clinically discordant results across chemistry platforms. Methods Chemistry testing was performed on Beckman-Coulter AU5800 series and Abbott Alinity C analyzers. Unexpected results prompted repeat testing, cross-platform comparison, review of reagent lot information, and correlation with clinical findings. LDH concentrations were evaluated in the context of assay methodology, and in one case, a reagent recall. Results Case one presented a patient with fulminant hepatic injury and markedly elevated LDH, from ∼4000 U/L to ∼9000 U/L, and AST (5,000–13,000) and ALT (∼3,000–6,000 U/L) increasing over the course of 10 hr. Serum bicarbonate measured on the AU5800 analyzer was unexpectedly elevated, corresponding to increases in LDH, producing a negative anion gap inconsistent with the clinical presentation. Ethanol testing on the AU5800, which utilizes NAD/NADH-coupled enzymatic methodology, was negative. Repeat testing on the Abbott Alinity C demonstrated substantially lower bicarbonate values consistent with the patient’s acid-base status. Review identified a Beckman-Coulter recall affecting certain bicarbonate reagent lots in samples with elevated LDH (one lot impacted at LDH >600 U/L; additional lots at LDH >2,000 U/L). As LDH decreased to only mildly elevated levels (300 U/L), bicarbonate values normalized on the Beckman platform and agreed with the Abbott Alinity. Case two involved a patient with septic shock, profound tissue injury, and markedly elevated AST (∼40,000 U/L) with ALT below the analytical measurement range, a pattern discordant with the clinical picture. There was extreme LDH elevation (>15,000–17,000 U/L). Ethanol measured on the Abbott Alinity platform was falsely positive (∼15 mg/dL), and bicarbonate was decreased (11 mmol/L). Given the assay’s NAD/NADH-coupled enzymatic methodology, interference from considerably elevated endogenous LDH was suspected. Conclusion These cases demonstrate that extreme LDH elevation can interfere with NAD/NADH-dependent chemistry assays, resulting in discrepant values and misinterpretations. The degree of interference may vary by platform and reagent lot. Although some assay formulations may include LDH inhibitors, such as sodium oxamate, interference at pathologic LDH concentrations may not be fully characterized or labeled in manufacturer documentation. Laboratories should consider LDH-mediated interference when results are discordant with the clinical presentation. Recognition of this mechanism and cross-platform verification when appropriate are essential to ensure accurate interpretation and patient safety.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 48 Bicarbonate Measurement Discrepancies Between Blood Gas and Chemistry Analyzers: Role of Reagent Variability and LDH InterferenceAmerican Journal of Clinical Pathology · 2026
- B-013 Severe Discordance Between Enzymatically Measured tCO2 and Calculated Bicarbonate with Extreme Lactate Dehydrogenase and Lactic AcidosisClinical Chemistry · 2026
- Pseudo-Metabolic Alkalosis: A Hidden Laboratory Pitfall With Elevated LDH.The Yale journal of biology and medicine · 2026
- 59 Superficially Elevated Serum Bicarbonate Level in a Critically Ill Patient: A Case of Critical Analytical InterferenceAmerican Journal of Clinical Pathology · 2026
- A-214 Hemolysis Interference for ALT and AST with Pyridoxal Phosphate Activation on 2 Different Roche Chemistry AnalyzersClinical Chemistry · 2026
- Rethinking the Limit of Detection for Serum Ethanol: Evidence from a Clinical Laboratory.EJIFCC · 2026