48 Bicarbonate Measurement Discrepancies Between Blood Gas and Chemistry Analyzers: Role of Reagent Variability and LDH Interference

Allison Martin, Richard Revia, Melkon Dombourian, Adekunle Alabi

Observational · n=413

Key points

From the paper's conclusion
  • These findings demonstrate that bicarbonate measurement discrepancies between blood gas and chemistry analyzers occur with clinically meaningful frequency and implicate reagent-related factors, including potential lot-to-lot variability and suboptimal inhibition of LDH-mediated interference, as significant contributors beyond elevated serum LDH alone; systematic evaluation of reagent integrity and

Abstract

This study investigated bicarbonate measurement discrepancies between blood gas analyzers and chemistry analyzers to determine the frequency, magnitude, and potential causes of instrumental bias in critically ill patients. A retrospective analysis of paired bicarbonate measurements from 413 patients between July 2024 and December 2024 was conducted using blood gas analyzers (ABL90 FLEX) and chemistry analyzers (Beckman Coulter AU5800), with serum lactate dehydrogenase (LDH) concentrations evaluated in discordant cases to assess for LDH-mediated interference in the enzymatic bicarbonate assay. Fifty-four instances of significant relative instrumental bias were identified, with serum LDH data available for 16 of these discordant cases; notably, bicarbonate discordance clustered temporally following the introduction of a new reagent lot, and patients with severe tissue injury exhibited the highest frequency of discordant values. These findings demonstrate that bicarbonate measurement discrepancies between blood gas and chemistry analyzers occur with clinically meaningful frequency and implicate reagent-related factors, including potential lot-to-lot variability and suboptimal inhibition of LDH-mediated interference, as significant contributors beyond elevated serum LDH alone; systematic evaluation of reagent integrity and analyte-specific interference patterns is warranted to improve bicarbonate measurement reliability in enzymatic assays used by chemistry analyzers.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers