48 Bicarbonate Measurement Discrepancies Between Blood Gas and Chemistry Analyzers: Role of Reagent Variability and LDH Interference
Allison Martin, Richard Revia, Melkon Dombourian, Adekunle Alabi
Key points
From the paper's conclusion- These findings demonstrate that bicarbonate measurement discrepancies between blood gas and chemistry analyzers occur with clinically meaningful frequency and implicate reagent-related factors, including potential lot-to-lot variability and suboptimal inhibition of LDH-mediated interference, as significant contributors beyond elevated serum LDH alone; systematic evaluation of reagent integrity and
Abstract
This study investigated bicarbonate measurement discrepancies between blood gas analyzers and chemistry analyzers to determine the frequency, magnitude, and potential causes of instrumental bias in critically ill patients. A retrospective analysis of paired bicarbonate measurements from 413 patients between July 2024 and December 2024 was conducted using blood gas analyzers (ABL90 FLEX) and chemistry analyzers (Beckman Coulter AU5800), with serum lactate dehydrogenase (LDH) concentrations evaluated in discordant cases to assess for LDH-mediated interference in the enzymatic bicarbonate assay. Fifty-four instances of significant relative instrumental bias were identified, with serum LDH data available for 16 of these discordant cases; notably, bicarbonate discordance clustered temporally following the introduction of a new reagent lot, and patients with severe tissue injury exhibited the highest frequency of discordant values. These findings demonstrate that bicarbonate measurement discrepancies between blood gas and chemistry analyzers occur with clinically meaningful frequency and implicate reagent-related factors, including potential lot-to-lot variability and suboptimal inhibition of LDH-mediated interference, as significant contributors beyond elevated serum LDH alone; systematic evaluation of reagent integrity and analyte-specific interference patterns is warranted to improve bicarbonate measurement reliability in enzymatic assays used by chemistry analyzers.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-013 Severe Discordance Between Enzymatically Measured tCO2 and Calculated Bicarbonate with Extreme Lactate Dehydrogenase and Lactic AcidosisClinical Chemistry · 2026
- 59 Superficially Elevated Serum Bicarbonate Level in a Critically Ill Patient: A Case of Critical Analytical InterferenceAmerican Journal of Clinical Pathology · 2026
- B-278 Markedly Elevated Lactate Dehydrogenase as an Under-Recognized Interference in Enzymatic Chemistry Assays Utilizing NAD/NADH-Coupled ReactionsClinical Chemistry · 2026
- Spectrum of Preanalytical Laboratory Errors in Clinical Biochemistry Laboratory in a Tertiary Care Center in Nepal.Kathmandu University medical journal (KUMJ) · 2025
- B-346 Storage Stability of HbA1c in Whole Blood: Method-Dependent Interference on Ion-Exchange HPLC and Degradation Kinetics Characterized by LC–MSClinical Chemistry · 2026
- Impact of Hemolysis on UCH-L1 Quantification in Serum and Plasma.EJIFCC · 2026