A-214 Hemolysis Interference for ALT and AST with Pyridoxal Phosphate Activation on 2 Different Roche Chemistry Analyzers
Adam McShane, Jessica Zakrajsek, Jennifer Zakrajsek
Key points
From the paper's conclusion- Methods The hemolysis interference studies were performed on the cobas c702 and cobas c503 (Roche Diagnostics) chemistry analyzers with pyridoxal phosphate activated reagents (c702: ALTPM and ASTPM; c503 ALTP2 and ASTP2).
Abstract
Background In vitro hemolysis is a common endogenous interference encountered in the clinical laboratory. Two common hepatic markers impacted by hemolysis are alanine aminotransferase (ALT) and aspartate aminotransferase (AST). AST, for example, is found in red blood cells. Therefore, an increase in hemolysis will cause falsely increased serum AST results. To avoid releasing erroneous results, laboratories must assess hemolysis levels for acceptability. Two common chemistry analyzers, cobas c702 and cobas c503 (Roche Diagnostics), have differing H index thresholds despite being the same methodology. Further, both H indices thresholds are low for AST (20 or 25, no/slightly hemolyzed), which may lead to frequent cancellations and sample re-draws. To assess the impact of hemolysis on AST and ALT and determine appropriate thresholds, interference studies were conducted on both analyzers. Methods The hemolysis interference studies were performed on the cobas c702 and cobas c503 (Roche Diagnostics) chemistry analyzers with pyridoxal phosphate activated reagents (c702: ALTPM and ASTPM; c503 ALTP2 and ASTP2). Low, mid, and high ALT and AST pools, at approximately 50, 200, and 600 U/L, were divided into target and control pools. The target pools were spiked with hemolysate to achieve an H index of approximately 1000 (1000 mg/dL of hemoglobin or grossly hemolyzed) for ALT and 250 (250 mg/dL of hemoglobin or moderately hemolyzed) for AST. The target samples were then analyzed for ALT or AST and H index. An equal amount of saline, as the added hemolysate to the target pool, was added to the control pool. The control pool was then utilized to dilute the target pool as outlined in Clinical & Laboratory Standards Institute (CLSI) document C56. Each pool was measured for ALT or AST and H index. Hemolysis interference was calculated by the variance of the target pools to the non-hemolyzed control pool for ALT and AST. Interferograms were constructed for the low, and high ALT and AST samples. Results At the highest H index (H = 1000) tested for ALT, the low pool experienced the largest interference with an increase of 10% and 64% for the c702 and c503, respectively. Instrument kinetic alarms for ALT were found at all 3 concentrations tested and were at the lowest hemolysis threshold for the low pool (H index = 250) on the c503. For AST at the highest H index (H = 250), the low pool was also the most impacted with an increase of 77% and 70% for the c702 and c503, respectively. Using an acceptability criteria of 15%, the highest acceptable H index threshold for all 3 ALT concentrations was 200 for the c503 and 500 for the c702. For AST, the highest acceptable H index for all 3 concentrations was 70 for the c503 and c702. Conclusion Hemolyzed specimens will falsely elevate ALT and AST levels. This study found higher acceptable H index thresholds than listed by the vendor. Further the concentrations tested experienced differential interference allowing for tiered-based hemolysis thresholds. These interventions would reduce unnecessary cancellations from in vitro hemolysis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optimization of hemolysis index thresholds for biochemical and immunoassays.Bioanalysis · 2026
- Assessment of Hemolysis, Icterus, and Lipemia Interference in 14 Clinical Chemistry and 5 Immunoassay Parameters.Clinical laboratory · 2026
- A-367 Assessment of Hemolysis Interference on HbA1c Measurement Using the Siemens Healthineers DCA Vantage AnalyzerClinical Chemistry · 2026
- A-204 Chemical modification of the hemolysis index with methylene blue on an automated chemistry analyzer: The holy grail?Clinical Chemistry · 2026
- B-011 Evaluation of the Tosoh HLC-723GR01 HbA1c Analyzer Using Normal Hemoglobin Samples and in the Presence of Common Hemoglobin and Thalassemic VariantsClinical Chemistry · 2026
- When Hemolysis Misleads: Implications for the Clinical Interpretation of Enzyme Activities in Pediatric Samples.Medicina (Kaunas, Lithuania) · 2026