Rethinking the Limit of Detection for Serum Ethanol: Evidence from a Clinical Laboratory.
José Arnau Pulido, Alejandra Pérez, Yolanda Fernández-Cladera, María Doladé, Susana Malumbres, Carla Fernández-Prendes
Laboratory Medicine Department, Hospital Universitari Germans Trias i Pujol, Badalona, Spain.
Key points
From the paper's conclusion- We report four cases in our laboratory with positive ethanol results of 11-14 mg/dL that, after reprocessing, were reported as "not detected".
Abstract
Ethanol is a widely consumed psychoactive substance. Serum concentrations above 100 mg/dL can cause cognitive impairment and confusion, while levels exceeding 300 mg/dL may be fatal. Thus, rapid and accurate determination of blood ethanol concentration is crucial for diagnosing and managing acute intoxication. Ethanol measurement also has important legal implications, such as in traffic accidents or criminal proceedings. Gas Chromatography-Mass Spectrometry (GC-MS) is the reference method for ethanol determination due to its accuracy and specificity. In routine clinical laboratories, however, ethanol quantification is usually performed using enzymatic spectrophotometric assays. Alcohol dehydrogenase (ADH) oxidizes ethanol to acetaldehyde with NAD as coenzyme, which is simultaneously reduced to NADH. The increase in NADH absorbance at 340 nm allows indirect quantification of ethanol. Although analytical characterization and method optimization are the manufacturers' responsibilities, literature indicates these processes are not standardized, and specific procedures are generally not disclosed. Consequently, clinical laboratories may need to verify these values, particularly during implementation of new methods, internal validation, or when analytical results conflict with clinical presentation. We report four cases in our laboratory with positive ethanol results of 11-14 mg/dL that, after reprocessing, were reported as "not detected". This highlights the need to determine the smallest analyte quantity detectable with certainty, commonly expressed as the limit of detection (LoD).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Extended blood ethyl glucuronide persistence after controlled ethanol administration: Implications for ethanol-negative delayed DUI cases.Forensic science international · 2026
- Analytical evaluation of a urine immunoassay for 2-ethylidene-1,5-dimethyl-3,3-diphenyl pyrrolidine (EDDP) with focus on cutoff optimization.Journal of forensic sciences · 2026
- B-390 Phosphatidylethanol (PEth) 16/18.1 and 16/18.2 quantification in whole blood and dried blood spotsClinical Chemistry · 2026
- The limited predictive value of initial serum ethanol concentration for assessing risk of complicated alcohol withdrawal: a retrospective cohort analysis.Alcohol and alcoholism (Oxford, Oxfordshire) · 2026
- Can Post-Mortem Cartilage Reliably Preserve Ethanol? A Time-Course Comparison with Blood and Urine.Toxics · 2026
- B-278 Markedly Elevated Lactate Dehydrogenase as an Under-Recognized Interference in Enzymatic Chemistry Assays Utilizing NAD/NADH-Coupled ReactionsClinical Chemistry · 2026