59 Superficially Elevated Serum Bicarbonate Level in a Critically Ill Patient: A Case of Critical Analytical Interference
Arianna Rahimian, Leenah Abojaib, Ali Kashkouli
Key points
From the paper's conclusion- This case highlights a critical analytical interference in serum bicarbonate measurement caused by markedly elevated LDH levels, resulting in spurious metabolic alkalosis and pseudo-hypercarbia on chemistry panels.
Abstract
A 77 year old male with a past medical history of multiple myeloma presented to the emergency department with respiratory distress and was found to have extensive pulmonary emboli and gastrointestinal bleeding. The patient’s laboratory values were significant for profound lactic acidosis, requiring continuous renal replacement therapy (CRRT). Despite delivering post-filter bicarbonate through CRRT, the patient’s acidemia continued to worsen, with a maximum pH of 7.12, a lactate of 22.9 mEq/L, and a lactate dehydrogenase (LDH) of 3,600 mEq/L. However, the comprehensive metabolic panels (CMP) reported gradually rising bicarbonate levels, up to 39 mEq/L. This case highlights a critical analytical interference in serum bicarbonate measurement caused by markedly elevated LDH levels, resulting in spurious metabolic alkalosis and pseudo-hypercarbia on chemistry panels. The discrepancy between enzymatic bicarbonate assays and blood-gas derived bicarbonate underscores the importance of understanding assay methodology and recognizing conditions that compromise analytical accuracy. This case illustrates how careful correlation of laboratory methods, potential assay interferences, and consultation with pathologists is essential, particularly for critically ill patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pseudo-Metabolic Alkalosis: A Hidden Laboratory Pitfall With Elevated LDH.The Yale journal of biology and medicine · 2026
- 48 Bicarbonate Measurement Discrepancies Between Blood Gas and Chemistry Analyzers: Role of Reagent Variability and LDH InterferenceAmerican Journal of Clinical Pathology · 2026
- B-013 Severe Discordance Between Enzymatically Measured tCO2 and Calculated Bicarbonate with Extreme Lactate Dehydrogenase and Lactic AcidosisClinical Chemistry · 2026
- B-278 Markedly Elevated Lactate Dehydrogenase as an Under-Recognized Interference in Enzymatic Chemistry Assays Utilizing NAD/NADH-Coupled ReactionsClinical Chemistry · 2026
- Sodium Bicarbonate for Critically Ill Adults with Metabolic Acidosis and ShockNew England Journal of Medicine · 2026
- Severe Suspected Metformin-Associated Lactic Acidosis With Relatively Preserved Serum Creatinine at Presentation: Sequential Intermittent Hemodialysis and Continuous Venovenous Hemodialysis.Cureus · 2026