B-013 Severe Discordance Between Enzymatically Measured tCO2 and Calculated Bicarbonate with Extreme Lactate Dehydrogenase and Lactic Acidosis
Rongrong Huang, Tavat Buraruk, Shella Manar
Key points
From the paper's conclusion- Clinical laboratorians should treat large tCO2–HCO3? discordances as potential analytical interference after ruling out preanalytical causes, correlate with LDH (and extreme lactate states), and recommend confirmation by ABG-derived bicarbonate and/or an alternative CO2 methodology.
Abstract
Background Total carbon dioxide (tCO2) measured on chemistry analyzers is expected to approximate bicarbonate (HCO3-) calculated from arterial blood gas (ABG) results (pH and pCO2). Large discrepancies unexplained by acid–base status may indicate preanalytical error, specimen mix-up, or analytical interference. Enzymatic tCO2 assays commonly quantify NADH?NAD+ conversion via spectrophotometric measurement of NADH consumption and can be vulnerable to methodology-specific interferences. A recent Beckman Coulter recall reported up to ∼30% positive bias in tCO2 at lactate dehydrogenase (LDH) of 2,000 U/L. Here, we describe a case of persistent, clinically implausible divergence between enzymatic tCO2 and ABG-calculated HCO3? in the setting of profound metabolic derangement, highlighting a potential diagnostic pitfall for clinical laboratorians. Methods A critically ill patient with diffuse large B-cell lymphoma (DLBCL) was admitted for sepsis requiring intensive care. Serial tCO2 as part of basic metabolic panels (Beckman AU5800 chemistry analyzer) and HCO3? calculated from pH and pCO2 (ABL90 FLEX PLUS blood gas analyzer) were reviewed for concordance. Markers of tissue hypoperfusion and tumor burden/cell turnover (lactate and LDH) were assessed. Discordance was evaluated in the context of acid–base physiology using the Henderson–Hasselbalch relationship. In addition, a retrospective lookback (7/1/2025-10/15/2025) was conducted to assess the frequency of high LDH, degree of tCO2 discordance, and potential clinical impact on patient management. Results A persistent, physiologically discordant pattern was noted by the ICU clinicians: pH 7.2–7.3 with pCO2 appx. 40 mmHg and calculated HCO3? appx. 17 mmol/L, while tCO2 repeatedly measured around 45 mmol/L (>2× higher). Lactate was markedly elevated (7 mmol/L) and LDH was extreme (>6,000 U/L). The large discrepancy suggested possible reagent lot–specific susceptibility of LDH interference. In addition, profound LDH elevation with suspected increased pyruvate flux (type B/Warburg physiology in lymphoma and/or critical illness) was considered a plausible mechanistic contributor. In the lookback, a total of 2,022 LDH results from 856 patients were reviewed. LDH >600 U/L occurred in 188 results (9.3%) from 65 patients (7.6%); LDH >2,000 U/L occurred in 35 results (1.7%) from 13 patients (1.5%). Of 65 high-LDH patients, 18 were deceased and 2 records were inaccessible. Chart review was performed for all accessible patients with LDH >2,000 U/L (6/6) and for a subset with LDH 600–2,000 U/L (13/39). Markedly elevated tCO2 discordant with clinical acid–base status and HCO3? was observed in multiple cases; however, no clinically significant impact on patient management was identified from documentation review alone, likely reflecting reliance on serial ABGs in severely ill patients. Conclusion The presented case highlights a clinically important pitfall that NADH-coupled enzymatic tCO2 methods can yield clinically misleading results when extreme metabolic abnormalities are present (e.g., profound LDH elevation), producing values incompatible with HCO3? and the observed pH/pCO2. Clinical laboratorians should treat large tCO2–HCO3? discordances as potential analytical interference after ruling out preanalytical causes, correlate with LDH (and extreme lactate states), and recommend confirmation by ABG-derived bicarbonate and/or an alternative CO2 methodology. Proactive communication of this pitfall with clinicians may prevent erroneous acid–base interpretation and inappropriate clinical interventions in critically ill patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 48 Bicarbonate Measurement Discrepancies Between Blood Gas and Chemistry Analyzers: Role of Reagent Variability and LDH InterferenceAmerican Journal of Clinical Pathology · 2026
- B-278 Markedly Elevated Lactate Dehydrogenase as an Under-Recognized Interference in Enzymatic Chemistry Assays Utilizing NAD/NADH-Coupled ReactionsClinical Chemistry · 2026
- 59 Superficially Elevated Serum Bicarbonate Level in a Critically Ill Patient: A Case of Critical Analytical InterferenceAmerican Journal of Clinical Pathology · 2026
- Pseudo-Metabolic Alkalosis: A Hidden Laboratory Pitfall With Elevated LDH.The Yale journal of biology and medicine · 2026
- Time-of-Day-Associated Variation in Serum CO2 Patient-Result Distribution and Its Impact on PBRTQC Performance.Diagnostics (Basel, Switzerland) · 2026
- The lactate-to-actual-bicarbonate ratio for risk stratification in adult patients with sepsis: a retrospective cohort studyFrontiers in Medicine · 2026