B-136 Transition of Freelite® assays from CobasTM 6000 c501 platform to CobasTM Pro c503 and CobasTM Pure c303

Robert van Heezik, Luke Moore, Jake Fisher, Rhiannon Stanton, Hannah Jones, Valerija Aleksejeva, Sephora Mbuyi, Felix Kwok,

Key points

From the paper's conclusion
  • This study describes adaptation of the Freelite® Kappa free and Lambda free assays from the c501 to the c303 and c503 platforms and evaluates whether analytical performance characteristics are maintained.

Abstract

Background Freelite® assays are currently available on the CobasTM 6000 c501 (Roche Diagnostics), which is approaching end of life. Laboratories are transitioning to the CobasTM Pro c503 or CobasTM Pure c303 analysers. This study describes adaptation of the Freelite® Kappa free and Lambda free assays from the c501 to the c303 and c503 platforms and evaluates whether analytical performance characteristics are maintained. Regulatory submissions will be completed and applicable regulatory approvals obtained to support commercial launch of the adapted assays. Methods Instrument capability differences between the CobasTM c501 and the c303 and c503 were assessed to recreate c501 assay parameters on the new platforms. Method comparison, onboard stability, analytical sensitivity, linearity, and precision studies were performed. Method comparison included 134 serum samples from healthy donors and patients with monoclonal gammopathies. Passing-Bablok regression and Bland-Altman analyses were performed. Onboard stability was evaluated over 106 days using three serum levels (normal and depleted serum), tested six times per time point alongside kit controls. Analytical sensitivity (limit of blank, limit of detection, and limit of quantification) was assessed over 3 days using four samples per level, tested five times per day. Linearity was evaluated using high and low serum pools spanning 10% above and below the measuring range, blended in 10% increments and tested in quadruplicate. Precision was assessed over 20 days using five serum samples spanning the assay range, tested in duplicate twice daily with >2 hours between runs. Results Both assays were successfully implemented using Roche CDC file creator software, with reaction proportions maintained within ±0.1% of the c501 configuration. Passing-Bablok regression demonstrated strong agreement between platforms. For c303 versus c501, slopes were 0.984 (Kappa; range 1.73–2408 mg/L) and 0.965 (Lambda; range 0.76–1196 mg/L). For c503 versus c501, slopes were 0.998 (Kappa) and 0.940 (Lambda). Bland-Altman analysis showed mean biases of -6.7% (Kappa) and -5.7% (Lambda) for c303, and -7.1% (Kappa) and -8.3% (Lambda) for c503. Onboard stability was maintained, with both assays achieving 94 days onboard on both instruments. Following 106 days, measurand differences were <2.3% from day 0. Analytical sensitivity was confirmed at 0.8 mg/L (Kappa) and 0.7 mg/L (Lambda) on both platforms, with CV <14.1%. Assays were linear from 3.25–61.63 mg/L (Kappa) and 5.16–85.45 mg/L (Lambda), with non-linearity <3% for Kappa and <6% for Lambda. Total CV for precision was 5.4% (c303) and 3.4% (c503) for Kappa, and 3.8% (c303) and 5.1% (c503) for Lambda. Conclusion Freelite® Kappa free and Lambda free assays were successfully transitioned from the CobasTM c501 to the CobasTM c303 and c503 platforms while maintaining analytical performance characteristics, including sample agreement, and precision.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers