B-112 Method Comparison of the Binding Site and Diazyme Serum Free Light Chain Assays on the Roche cobas c502

Spencer Seely, Megan Palacio, Michael Kelner, Robert Fitzgerald, Raymond Suhandynata

Key points

From the paper's conclusion
  • This study evaluated the analytical performance of the Diazyme Human Kappa and Lambda Free Light Chain assays relative to the Binding Site Freelite assay when performed on the Roche cobas c502 automated chemistry analyzer.

Abstract

Background Monoclonal gammopathies, including multiple myeloma, are characterized by clonal proliferation of immunoglobulin-producing plasma cells. Excess production of serum free light chains (sFLCs) may result in renal injury, amyloidosis, and other end-organ damage. Accurate quantification of kappa and lambda sFLCs is essential for diagnosis, disease monitoring, and therapeutic response assessment in patients with multiple myeloma. The Binding Site Freelite assay is widely used for sFLC measurement; however, high analyte concentrations combined with on-board instrument dilutions introduce an increased risk of error propagation, particularly in the presence of systematic bias arising from sample carryover on automated chemistry platforms. This study evaluated the analytical performance of the Diazyme Human Kappa and Lambda Free Light Chain assays relative to the Binding Site Freelite assay when performed on the Roche cobas c502 automated chemistry analyzer. Methods A total of 140 de-identified patient plasma samples were analyzed on the Roche cobas c502 using both Freelite and Diazyme FLC assays. Reproducibility was evaluated by testing 3 pooled samples (10 mg/L, 50 mg/L, and >100 mg/L) repeated over 10 days. Linearity was assessed by serial dilution in triplicate. Remnant CAP proficiency testing material was tested on both platforms in duplicate. Descriptive statistics (mean, median, SD, IQR, and %CV) were calculated for kappa, lambda, and kappa:lambda ratios. Statistical analyses included correlation analysis, percent bias assessment, and paired Wilcoxon signed-rank testing. Results Patient results demonstrated a right-skewed distribution with extreme values approaching the upper analytical measurement range. Diazyme and Freelite assays showed moderate correlations for kappa (R² = 0.786) and lambda (R² = 0.842). Correlation of the kappa:lambda ratio between assays was poor (R² = 0.549). Both assays demonstrated acceptable reproducibility, with %CVs < 5%. Diazyme kappa FLC concentrations exhibited negative bias, whereas lambda FLC concentrations showed positive bias relative to Freelite. These biases were consistent across analysis of CAP proficiency testing material Conclusion High sFLC concentrations may disproportionately influence clinical interpretation, particularly for patients undergoing longitudinal monitoring and transitioning between assays, necessitating reestablishment of assay-specific baseline concentrations prior to method transition.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

B-112 Method Comparison of the Binding Site and Diazyme Serum Free Light Chain Assays on the Roche cobas c502 — Science Skim