B-114 Evaluation of the Diazyme Serum Free Light Chain Assay on the Abbott Alinity c Platform: Comparison with Freelite Assay and Reference Interval Concordance
Yasser Tabana, Casey Williamson, Mark Griffiths, Sophie Furlow, Randal Schneider, David Daghfal, Vathany Kulasingam
Key points
From the paper's conclusion- Therefore, we aimed to evaluate analytical concordance between the Diazyme sFLC assay performed on the Abbott Alinity c platform and the Freelite assay performed on the Optilite analyzer Methods Serum samples from 297 patients were analyzed for kappa, lambda, and the k/l ratio using either Diazyme’s kappa and lambda assays on the Abbott Alinity c platform, or Freelite assays on the Optilite analyz
Abstract
Background Serum free light chains (sFLC) are essential for the diagnosis and monitoring of plasma cell dyscrasias. However, clinical guidelines are largely based on Freelite (The Binding Site) sFLC assays, creating challenges when alternative methods are implemented. Therefore, we aimed to evaluate analytical concordance between the Diazyme sFLC assay performed on the Abbott Alinity c platform and the Freelite assay performed on the Optilite analyzer Methods Serum samples from 297 patients were analyzed for kappa, lambda, and the k/l ratio using either Diazyme’s kappa and lambda assays on the Abbott Alinity c platform, or Freelite assays on the Optilite analyzer. Method comparisons were performed using Passing-Bablok regression and Pearson correlation. Results were classified as low, normal, or high using either manufacturer-specific reference intervals (RIs) or iSTOPMM RIs for individuals above or below 70 years of age. Concordance between platforms was assessed with simple percent agreement and Cohen’s kappa Results Linear relationships and correlations were observed between platforms. Pearson correlation coefficients for kappa FLC, lambda FLC, and the k/l ratio were 0.941, 0.690, and 0.884, respectively (all P<0.0001). Using manufacturer RIs, concordance was highest for kappa FLC and lambda FLC (85% and 83%; Cohen’s kappa =0.704-0.706), while k/l ratio classifications showed only moderate concordance (76%). Application of iSTOPMM RIs resulted in high concordance across all analytes (kappa FLC =82%, lambda FLC =90%, k/l =83%) Conclusion The Diazyme sFLC assay demonstrates good agreement with the Freelite assay when interpreted using manufacturer-specific or iSTOPMM RIs. Further studies should determine whether residual discordances have clinically meaningful implications
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-135 Evaluation of Serum Free Light Chains: A Method Comparison Between the Diazyme Assay on the Beckman-Coulter AU5800 and the Binding Site Assay on the Optilite PlatformClinical Chemistry · 2026
- B-137 Diazyme and Freelite Serum Free Light Chain Assays Are Not Clinically Interchangeable at IMWG-Defined Decision PointsClinical Chemistry · 2026
- Consistency and correlation of a domestic Kehua serum free light chain assay with two imported analytical systems.Frontiers in oncology · 2026
- B-112 Method Comparison of the Binding Site and Diazyme Serum Free Light Chain Assays on the Roche cobas c502Clinical Chemistry · 2026
- Noninterchangeability of Serum-Free Light Chain Assays at IMWG Thresholds: Analysis of 2117 Paired Samples.Clinical lymphoma, myeloma & leukemia · 2026
- B-197 The abnormality of “normal”: performance and verification of serum free light chain ratio reference intervals in donor sera populationsClinical Chemistry · 2026