B-135 Evaluation of Serum Free Light Chains: A Method Comparison Between the Diazyme Assay on the Beckman-Coulter AU5800 and the Binding Site Assay on the Optilite Platform
Annalara Fischer, Karen Cost, Cierra Sharp, Nicole Pozzi, Morgan McCoy, Mohamed Hegazi, Mustafa Kawaaz, Saeed Jortani
Key points
From the paper's conclusion- This study aimed to compare the analytical performance of the Diazyme serum FLC assay on the Beckman-Coulter AU5800 analyzer with the Binding Site assay on the Optilite® platform.
Abstract
Background Serum free light chains (FLCs) are essential biomarkers for the diagnosis, prognosis, and monitoring of plasma cell disorders, such as multiple myeloma. Clinical interpretation of these values is guided by the International Myeloma Working Group (IMWG) diagnostic and response criteria, which were established exclusively using the historically predominant The Binding Site FLC (Freelite®) assay. As alternative FDA-cleared assays, such as the Diazyme Laboratories FLC assay, are deployed on high-throughput chemistry analyzers and increasingly adopted by clinical laboratories, it is critical to understand inter-assay differences and their potential impact on clinical interpretation and longitudinal tracking. This study aimed to compare the analytical performance of the Diazyme serum FLC assay on the Beckman-Coulter AU5800 analyzer with the Binding Site assay on the Optilite® platform. Methods Remnant sera submitted in October 2025 for free light chain testing as a part of routine patient care were analyzed using both the Diazyme and Binding Site assays. Kappa (?) and lambda (?) FLC concentrations, as well as ?/? ratios, were compared across the measured ranges of the study (?FLC: 3.50 – 684.00 mg/L, n=38; ?FLC: 1.74 – 16,782.60 mg/L, n=39). Method comparison analyses were performed to evaluate bias, and analyzed using EP Evaluator®. Additional dilution/linearity and spike-recovery studies were performed on select discrepant samples. Results The Diazyme ?FLC assay demonstrated a significant positive proportional bias relative to the Binding Site assay with a correlation coefficient of 0.9909, slope of 2.887, and intercept of -39.420. The Diazyme ?FLC assay did not demonstrate substantial proportional deviation by regression analysis (slope= 0.807, intercept= 9.723), but rather exhibited a correlation coefficient 0.8417 with several select patient samples (>120 mg/L) exhibiting considerable negative bias with the Diazyme assay. Dilution/linearity or spike-recovery studies performed on three discrepant ?FLC demonstrated appropriate linearity using Diazyme, suggesting a potential issue of differential antibody reactivity to patient-specific monoclonal proteins resulting in negative interference in select cases. The downstream ?/? ratios were also impacted by this discrepancy, with a correlation coefficient of 0.9031, slope of 0.423, and an intercept of 0.691 (n=36). Conclusion Significant analytical differences exist between the Diazyme and Binding Site serum FLC assays, highlighting that results are not interchangeable. Because current IMWG diagnostic and response criteria were established from studies solely using the Binding Site assay, consideration should be given when implementing other FDA-cleared methods to support effective patient care. Laboratories implementing alternative FLC assays must perform robust assay-specific validation, including method comparison with the guideline-derived Binding Site assay and reportable range verification, and ensure clinician education and clear communication to prevent misinterpretation. We further recommend harmonization and international standardization of serum FLC assays to promote consistent clinical interpretation and effective patient care.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-137 Diazyme and Freelite Serum Free Light Chain Assays Are Not Clinically Interchangeable at IMWG-Defined Decision PointsClinical Chemistry · 2026
- B-114 Evaluation of the Diazyme Serum Free Light Chain Assay on the Abbott Alinity c Platform: Comparison with Freelite Assay and Reference Interval ConcordanceClinical Chemistry · 2026
- B-112 Method Comparison of the Binding Site and Diazyme Serum Free Light Chain Assays on the Roche cobas c502Clinical Chemistry · 2026
- Consistency and correlation of a domestic Kehua serum free light chain assay with two imported analytical systems.Frontiers in oncology · 2026
- Noninterchangeability of Serum-Free Light Chain Assays at IMWG Thresholds: Analysis of 2117 Paired Samples.Clinical lymphoma, myeloma & leukemia · 2026
- B-189 Comparative Analysis of Free Light Chain Assay and Electrophoretic Techniques in the Laboratory Assessment of Monoclonal AbnormalitiesClinical Chemistry · 2026