Consistency and correlation of a domestic Kehua serum free light chain assay with two imported analytical systems.

Yansong Ren, Shoulei Wang, Limin Han, Yansheng Wang, Qian Ren, Mengxin Wang, Yonghui Xia, Guoqing Zhu

State Key Laboratory of Experimental Hematology, National Clinical Research Center for Blood Diseases, Haihe Laboratory of Cell Ecosystem, Institute of Hematology & Blood Diseases Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Tianjin, China. · Shanghai Kehua Bio-Engineering Co., Ltd, Reagent R and D Center, Shanghai, China.

Key points

From the paper's conclusion
  • We evaluated a domestic latex-enhanced immunoturbidimetric Kehua serum sFLC assay against Freelite and N Latex systems.

Abstract

BACKGROUND: Serum free light chain (sFLC) testing is central to the diagnosis and monitoring of plasma cell disorders, but assay availability and implementation can vary by laboratory setting. We evaluated a domestic latex-enhanced immunoturbidimetric Kehua serum sFLC assay against Freelite and N Latex systems. METHODS: Residual serum samples from 133 adults with suspected or established monoclonal gammopathies were measured in parallel by Kehua and Freelite; 122 were also tested by N Latex. Precision was assessed with low- and high-level serum controls. Method agreement was evaluated using correlation, Passing-Bablok regression, Bland-Altman analysis, qualitative concordance and comparator-defined ROC analysis. RESULTS: Kehua within-run CVs were <5% for κ and λ controls. Agreement with Freelite was excellent for FLC-κ, FLC-λ and κ/λ ratio (Pearson r = 0.988-1.000; CCC = 0.983-0.999), with Passing-Bablok slopes close to unity. Agreement with N Latex was also strong (r = 0.991-0.998), although κ/λ ratio showed proportional bias. Total qualitative concordance was 89.47-93.23% versus Freelite and 83.61-91.80% versus N Latex, with Cohen's κ of 0.67-0.85. Comparator-defined ROC AUCs were 0.994-0.999 for elevated FLC-κ or FLC-λ and 0.796-0.869 for κ/λ ratio abnormality. CONCLUSION: The Kehua serum assay demonstrated acceptable precision and good clinical agreement with two imported methods. It may serve as a domestic option for routine serum sFLC quantification, provided method-specific intervals are maintained and the same platform is used for serial monitoring.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers