B-027 Development of a Hemolysis Interference Detection and Signal Separation Model for Pediatric Biochemical Testing Based on a Three-Dimensional PDR Platform to Enable Precision Diagnosis in Pediatric Diseases
Yan Liu, Fengqing Cai, Gui Hao, Shuting Jiang, Jin Li, Yong Dai, Youlin Liu, Gang Yu,
Key points
From the paper's conclusion- Overall, the research results effectively reduce misjudgement caused by hemolysis interference and better align with clinicians’ expectations in pediatric diagnosis and treatment.
Abstract
Background Hemolysis remains a pervasive preanalytical challenge in pediatric biochemical testing, frequently introducing significant analytical interference. This study develops an automated algorithm using the BS2800M Panorama-Dynamic-Recognition (PDR) platform to characterize and subtract hemolysisrelated signals from reaction mixtures, enabling more accurate results. Realworld clinical validation will assess its performance and applicability. Methods The PDR technology captures realtime 33cycle absorbance data and fullspectrum signals at 16 wavelengths, generating a threedimensional reaction spectrogram that enables algorithm development. The algorithm identifies hemolysis-specific spectral features, quantifies interference intensity using spectral interference coefficients (SIC), efficiently separates hemolysis-derived signals from target analyte signals through spectral deconvolution, and reconstructs interference-free reaction curves to generate corrected results. A hospitalwide survey assessed pediatric hemolysis status and perceptions, and 14 hemolysissusceptible biochemical parameters were selected. The corrected results underwent systematic validation—precision, accuracy, linearity, and clinical applicability—to evaluate their diagnostic value. Results This study defined a hemolysis index (HI) > 50 as hemolytic samples and analyzed 53,578 pediatric biochemical specimens across wards, disease categories, and age groups. Hospitalized children showed a markedly higher hemolysis rate than outpatients (34% vs 2%, P<0.05). The degree of hemolysis among inpatients was mainly distributed within HI 50–600 mg/dL, with a significantly higher proportion of mod erate hemolysis compared with severe hemolysis (30% vs 0.69%, P<0.05). Hemolysis rates exceeded 10% across all inpatient age groups, and children under three demonstrated the highest incidence. Patients with skin injuries (surface wounds, burns, scalds, lumps) and respiratory infections exhibited hemolysis rates above 40%, while several other disease categories—such as ENT disorders, urinary diseases, dermatological conditions, non-infectious respiratory diseases, and orthopedic conditions—also showed hemolysis rates exceeding 30%. Compared with true values, the new-algorithm–corrected “research results” demonstrated significantly improved accuracy and a higher proportion of measurements within allowable deviation limits (P<0.05). Precision was favorable, with CV% = 1/4TEa in more than 80% of samples, showing a typical “power-function” pattern where variation increased at lower concentrations, and precision did not differ across hemolysis levels (P>0.05). The research results of all 14 biochemical analytes exhibited excellent linearity (R²>0.999) within clinically relevant ranges and maintained appropriate reference-interval compliance. For liver disease screening and differential diagnosis, the AUC values of AST, ADA, and related markers were markedly improved in research results. Similar improvements were observed in myocardial disease screening. False-positive rates for AST, CK, CK-MB, and LDH in children without hepatic or cardiac diseases were significantly reduced (P<0.05). Overall, the research results effectively reduce misjudgement caused by hemolysis interference and better align with clinicians’ expectations in pediatric diagnosis and treatment. Conclusion The PDR-based algorithm effectively solves a long-standing industry challenge in pediatric biochemical testing by accurately recognizing and correcting hemolysis interference. It significantly reduces sample rejection and re-sampling, improves the reliability of liver and cardiac disease assessment—especially in emergency care—and enhances clinical decision-making. By improving data quality and workflow efficiency, the algorithm provides meaningful clinical and social value.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- When Hemolysis Misleads: Implications for the Clinical Interpretation of Enzyme Activities in Pediatric Samples.Medicina (Kaunas, Lithuania) · 2026
- Preanalytical interventions to reduce hemolysis and improve specimen quality: a systematic review and meta-analysis.Laboratory medicine · 2026
- B-273 Study on the Application Effect of a Novel Serum Index Detection Method Based on Reagent BackgroundClinical Chemistry · 2026
- Why do hemolysis interference studies produce different results? A proposal for standardized experimental design and reportingClinical Chemistry and Laboratory Medicine (CCLM) · 2026
- Simulation Study on Preanalytical Variables in Coagulation Testing: Interference Effects and Threshold Determination of Hemolysis, Lipemia, Icterus and Insufficient Sample Volume.Clinical and applied thrombosis/hemostasis : official journal of the International Academy of Clinical and Applied Thrombosis/Hemostasis · 2026
- A-277 Improving Sample Integrity: Preclinical Evidence Demonstrating Reduced Hemolysis with Novel Blood Collection DeviceClinical Chemistry · 2026