B-273 Study on the Application Effect of a Novel Serum Index Detection Method Based on Reagent Background

Puguang Chen, Hangmin Liu, Xuanmei Huang, Wenlong Lu, Helu Liu, Yingping Cao

Key points

From the paper's conclusion
  • Results On the Mindray BS-1000M automatic biochemical analyzer, a total of 10 biochemical test items, including ALT, AST, CK-MB, GLU, HCY, IGE, LDH, UA, UREA, and a-HBDH, could automatically output accurate RSI results during the detection process, which could reduce the serum index (SI) detection workload of about 85.6% of serum samples.

Abstract

Background Hemolysis, lipemia, and jaundice are the main causes of pre-analytical errors in laboratories, accounting for 40% to 60% of the sources of pre-analytical errors, which seriously affect the accuracy and reliability of test results. The traditional visual identification method has the disadvantages of low efficiency and poor consistency. Although the traditional normal saline serum index (NSI) detection can obtain reliable results, it has inherent defects such as occupying reaction cup positions, consuming normal saline, increasing labor intensity, and reducing detection efficiency. This study aims to evaluate the core performance indicators such as precision and accuracy of the new reagent-based serum index (RSI) detection scheme, verify whether it can effectively avoid the related problems of traditional NSI detection, and provide a more efficient and convenient serum index detection method for clinical testing. Methods A total of 315 serum samples, including normal serum samples and those with hemolysis, or jaundice interference, were collected. The precision, accuracy, and linear range of NSI and RSI were detected and evaluated on the Mindray BS-1000M automatic biochemical analyzer, respectively; meanwhile, methodological comparison was performed with three automatic biochemical analyzers, namely Roche C701, Beckman AU5800, and Ortho Vitros 5600. Typical real interference cases were used to verify the associated alarm function of RSI on the Mindray BS-1000M analyzer for interference in biochemical test items, and a comprehensive comparative analysis of NSI and RSI was conducted from the aspects of detection performance, practical functions, and operational ease of use. Results On the Mindray BS-1000M automatic biochemical analyzer, a total of 10 biochemical test items, including ALT, AST, CK-MB, GLU, HCY, IGE, LDH, UA, UREA, and a-HBDH, could automatically output accurate RSI results during the detection process, which could reduce the serum index (SI) detection workload of about 85.6% of serum samples. The intra-assay precision coefficient of variation (CV) of both NSI and RSI was < 1%, and there was no statistically significant difference between the two methods (P > 0.05); both detection methods had good linear relationships (r > 0.99) and correlations (R² > 0.99). The RSI of Mindray BS-1000M and the NSI of Roche C701 both showed good correlations with the theoretical concentrations of pure interferents (R² > 0.99 and R² > 0.98, respectively), while the NSI of Beckman AU5800 and Ortho Vitros 5600 had weak correlations with the theoretical concentrations of pure interferents due to the limitation of their detection upper limits. The verification results of typical real interference cases showed that RSI had a good associated alarm function for interference in biochemical items, which could effectively prompt detection interference. Conclusion The RSI detection scheme does not require additional consumption of normal saline, does not occupy separate reaction cup positions, can significantly reduce the frequency of manual intervention, solves the inherent defects of traditional NSI detection, and is an effective supplement to the traditional NSI detection scheme.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers