Simulation Study on Preanalytical Variables in Coagulation Testing: Interference Effects and Threshold Determination of Hemolysis, Lipemia, Icterus and Insufficient Sample Volume.
Xiang Zhang, Qing Liao, Hongwei Liu, Wenjuan Liu, Haiting Ye, Xiangdong Xu
Department of Laboratory Medicine, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China. · Werfen Medical Device Trading (Beijing) Co., Ltd, Beijing, China.
Key points
From the paper's conclusion- ObjectiveTo systematically evaluate the interference degree of hemolysis, lipemia, icterus, and sample collection volume on routine test items of the ACL TOP750 automatic coagulation analyzer, assess the technical performance of the HIL module, verify the manufacturer's anti-interference capability, and establish laboratory-specific anti-interference thresholds for optimizing preanalytical quality
Abstract
ObjectiveTo systematically evaluate the interference degree of hemolysis, lipemia, icterus, and sample collection volume on routine test items of the ACL TOP750 automatic coagulation analyzer, assess the technical performance of the HIL module, verify the manufacturer's anti-interference capability, and establish laboratory-specific anti-interference thresholds for optimizing preanalytical quality management.MethodsGradient interference models of hemolysis, icterus, and lipemia, as well as sample systems with different collection volumes, were constructed. Seven routine coagulation indicators (PT, aPTT, FIB, TT, DD, FDP, AT) were detected, with a relative bias (Bias%) of 10% set as the acceptable standard to comprehensively evaluate the instrument's anti-interference ability.ResultsThe H-index of ACL TOP750 was highly linearly correlated with hemoglobin concentration, the I-index with total bilirubin, and the L-index with triglyceride concentration (R2 ≥ 0.951). For hemolysis: TT bias was >10% at H ≥ 650 mg/dL, FIB bias was unacceptable at H ≥ 880 mg/dL, and DD bias was >10% at H ≥ 270 mg/dL. For lipemia, DD bias was >10% at L ≥ 1100 mAbs, and ultracentrifugation effectively improved lipemia interference. For icterus, aPTT bias was unacceptable at I ≥ 35 mg/dL, and PT and FIB biases were >10% at I ≥ 38 mg/dL. At a sample filling ratio of 70% (2.1 mL), biases of several analytes exceeded the acceptable limits; accordingly, 70% was defined as the threshold for sample rejection.ConclusionThe HIL module of the ACL TOP750 testing system exhibits excellent performance, and its anti-interference capability meets the manufacturer's specifications. The optimized HIL index and sample volume anti-interference thresholds established in this study are more in line with clinical practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assessment of Hemolysis, Icterus, and Lipemia Interference in 14 Clinical Chemistry and 5 Immunoassay Parameters.Clinical laboratory · 2026
- Reduction in Sample Recollection Number Following the Implementation of Automated Pre-Analytical Interference Detection in Coagulation Testing.Annals of clinical biochemistry · 2026
- Re-evaluating haemolysis rejection criteria for PT, aPTT, and D-dimer testing: evidence from paired clinical samples on the CN-3000 analyser.Advances in laboratory medicine · 2026
- Evaluation of Pre-Analytical Performance Using IFCC Quality Indicators, Six Sigma Metrics and Root Cause Analysis in a Biochemistry Laboratory-A Retrospective Study.Diagnostics (Basel, Switzerland) · 2026
- Spectrum of Preanalytical Laboratory Errors in Clinical Biochemistry Laboratory in a Tertiary Care Center in Nepal.Kathmandu University medical journal (KUMJ) · 2025
- 182 Leveraging an Automated Delta-Check Rule to Flag Critical Preanalytical Errors in Coagulation Testing: A 14-Month Lab Practice InterventionAmerican Journal of Clinical Pathology · 2026