Why do hemolysis interference studies produce different results? A proposal for standardized experimental design and reporting
Oswald Sonntag, Giuseppe Lippi, Charles R. Lefèvre, Hikmet Can Çubukçu, Mario Plebani, Nathan Lorde
Key points
From the paper's conclusion- Therefore, identical hemolysis indices do not necessarily represent equivalent experimental interferents.
Abstract
Background Hemolysis is a major preanalytical source of laboratory interference, but methods used to experimentally generate hemolyzed specimens are poorly standardized. Content Different lysis procedures, blood sources, cellular compositions, hemolysate volumes, control designs, and time-to-analysis can influence results independently of free hemoglobin concentration. The amount and composition of intracellular material released may also vary with erythrocyte, leukocyte, and platelet characteristics. Therefore, identical hemolysis indices do not necessarily represent equivalent experimental interferents. Summary Hemolysis interference studies should clearly distinguish the interferent, its method of generation, and its measurement. Detailed reporting of specimen preparation and experimental conditions is essential for reproducibility and comparison between studies. Outlook Harmonized hemolysate preparation and reporting, together with assessment at clinically relevant measurand concentrations and, when appropriate, multiple time points, would improve the reliability of hemolysis thresholds and laboratory decision-making.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Preanalytical interventions to reduce hemolysis and improve specimen quality: a systematic review and meta-analysis.Laboratory medicine · 2026
- When Hemolysis Misleads: Implications for the Clinical Interpretation of Enzyme Activities in Pediatric Samples.Medicina (Kaunas, Lithuania) · 2026
- Simulation Study on Preanalytical Variables in Coagulation Testing: Interference Effects and Threshold Determination of Hemolysis, Lipemia, Icterus and Insufficient Sample Volume.Clinical and applied thrombosis/hemostasis : official journal of the International Academy of Clinical and Applied Thrombosis/Hemostasis · 2026
- A-320 Verification of hemolysis levels to improve laboratory workflowClinical Chemistry · 2026
- A-325 Optimizing Management of Hemolyzed Specimens: Enhancing Clinical Value and Reducing Troponin and Potassium RecollectionsClinical Chemistry · 2026
- A-277 Improving Sample Integrity: Preclinical Evidence Demonstrating Reduced Hemolysis with Novel Blood Collection DeviceClinical Chemistry · 2026