A-186 Evaluation of Within-Tube Stability of BD Vacutainer® Rapid Serum Tubes for Select Chemistry Analytes and Cardiac Markers

Nasrin Al Thubian, Renee Howell, Charles Opara, Jeanne Li, Lingzhi Li

Key points

From the paper's conclusion
  • Conclusion These studies demonstrated that most chemistry analytes, cardiac markers, and ß-hCG remained stable in BD Vacutainer® Rapid Serum Tubes for up to 24 hours at room temperature, supporting the feasibility of add-on testing using previously collected samples.

Abstract

Background Add-on testing offers the potential to enhance the utilization of previously collected patient samples, thereby minimizing patient blood loss due to repeat collections and reducing the discomfort associated with multiple venipunctures. Additionally, it helps decrease waste and optimize time efficiency for both patients and laboratory personnel. Maintaining analyte stability is essential to guarantee accurate test results to support add-on testing. Establishing analyte stability within the laboratory is therefore fundamental for successful add-on testing. To address this, two prospective, multi-site randomized studies were conducted to evaluate within-tube analyte stability in BD Vacutainer® Rapid Serum Tubes (BD RST). Study 1 focused on selected chemistry analytes, while Study 2 assessed chosen cardiac markers and ß-hCG over a period of up to 24 hours at room temperature. Methods Blood samples were drawn into BD RST Tubes from 101 participants for Study 1, and 315 participants for Study 2. Each study included both inpatients and outpatients with cardiac disease, thyroid dysfunction, and other conditions likely to yield abnormal analyte results. In addition, contrived samples were created as needed to cover medically significant ranges for certain analytes. All tubes were handled and processed following the manufacturer’s recommended protocol. For each study, initial testing for all analytes was performed within 2 hours following collection. Samples were then kept in primary tubes at room temperature upright in a rack. For Study 1 testing was performed on the Beckman Coulter AU480 for Glucose at 18 and 24 hours, for CO2 at 12 and 18 hours, and other analytes were tested at 24 hours. For Study 2, testing was performed for select cardiac markers and ß-hCG at 24 hours on the Roche cobas® e411/e601. In each study, average differences between later time points vs. initial time with 95% confidence intervals were estimated and compared to the specified clinical acceptance limits (CAL) for each analyte to determine analyte stability. Results Study 1: Most chemistry analytes remained stable within tubes at 24 hours, except CO2, which was stable for 12 hours at room temperature. HDL showed a mean bias of 4.5% at 24 hours (within the 5% CAL), with only a minor upper confidence interval exceedance (by 0.4%), considered clinically acceptable given the reference range (23-92 mg/dL). Study 2: All tested cardiac markers and ß-hCG were stable in BD RST tubes for 24 hours. Conclusion These studies demonstrated that most chemistry analytes, cardiac markers, and ß-hCG remained stable in BD Vacutainer® Rapid Serum Tubes for up to 24 hours at room temperature, supporting the feasibility of add-on testing using previously collected samples. However, CO2 was not stable beyond 12 hours at room temperature, indicating that caution is needed when interpreting results for this analyte after delayed analysis. Overall, these findings provide valuable guidance for laboratories aiming to optimize efficiency and patient care through the implementation of add-on testing protocols.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A-186 Evaluation of Within-Tube Stability of BD Vacutainer® Rapid Serum Tubes for Select Chemistry Analytes and Cardiac Markers — Science Skim