Clinical Chemistry

B-282 Analytical Validation and Comparison of Becton Dickinson and Greiner Bio-One Blood Collection Tubes

Daniel Kirchhoff, Dipika Patel, Gonzalo Pinero, Juan Turati, Damodara Mendu

Key points

From the paper's conclusion
  • Despite several analytical differences, the comparison is acceptable, and Greiner Bio-One sample tubes are an acceptable alternative to BD sample tubes.

Abstract

Background In the summer of 2025, the Mount Sinai Department of Chemistry (MSH) received a request to evaluate the suitability of Greiner Bio-One (Greiner) branded blood collection tubes for use in patient testing. MSH historically has used Becton Dickinson (BD) branded tubes, and all analytical validations were performed using samples collected in BD tubes. This request is a preventative measure in case of supply chain disruptions with BD. Although sample types are identical (serum, plasma), this led to the question whether collection tube manufacturer could have any analytical effect or alter patient results. Assay comparative data provided by Greiner was dated and largely performed on the previous generation of chemistry analyzers (Abbott Architect). As such, MSH performed a large-scale study assessing the analytical performance of BD vs. Greiner collection tubes. Methods Blood was collected from 90 MSH employee volunteer donors. 4 Greiner vs. BD sample sets were collected, consisting of the following tube types: lithium heparin, sodium fluoride/potassium oxalate, EDTA, and serum separator tubes. As it's impossible to analyze all analytes measured at MSH; each sample set was assayed for a selection of 40 analytes deemed critical for patient care decisions. Analytes selected encompassed a wide range of methodologies, such as ISE, colorimetric, and immunoassay. Analytes were primarily measured on the Abbott Alinity CI, DiaSorin Liaison, Abbott Architect, and other platforms. A selection of analytes examined includes sodium, potassium, calcium, ALP, syphilis, CMV IgG, and hCG. Each Greiner and BD sample was assayed, and the resulting values were compared for congruency. Data was represented in terms of slope (S), correlation coefficient (r), and qualitative comparison, when appropriate. In addition, analyte stability was assessed at three temperature levels over seven days in all tube sets to determine if sample stabilities differed between the two tube manufacturers. Results The study afforded an extremely large data set. Notable results include: lithium heparin plasma: sodium, S: 0.817, r: 0.655; potassium, S: 0.967, r: 0.971; albumin, S: 1.021, r: 0.990; calcium, S: 0.853, r: 0.861; ALP S: 1.021, r: 0.91. Serum: sodium, S: 0.977, r: 0.964; potassium, S: 0.915, r: 0.843; albumin, S: 0.992, r: 0.980; calcium, S: 0.939, r: 0.945; ALP S: 1.003, r: 0.998; lithium, S: 1.031, r: 0.976, S: 1.115, r: 0.920; CMV IgG, S: 0.946, r: 0.984; hCG: 24/24 congruent samples. EDTA plasma: tacrolimus, S: 0.964, r: 0.994. Potassium oxalate/sodium fluoride plasma: lactate, S: 0.989, r: 0.922. Stability studies showed no significant differences in sample stabilities between the two manufacturers. Conclusion Sample collection is crucial to assay performance and accurate patient results. Provided identical sample types, analytical performance should be largely unaffected by the collection container manufacturer; however, there is limited published data exploring this comparison. Our studies show that Greiner samples tubes are largely analytically identical to BD branded tubes, with some exceptions in several plasma analytes. Sample stabilities are identical between the two manufacturers. Despite several analytical differences, the comparison is acceptable, and Greiner Bio-One sample tubes are an acceptable alternative to BD sample tubes.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers