Development of a 3D Electronic Detector Array for Radiotherapy Quality Assurance

Dante E. Roa, PhD, Miguel Risco-Castillo, Renzo Romero, Enzo Aucca, Renzo Ocampo, William de Souza Santos, PhD, Mirko Salomón Alva-Sánchez, PhD, Carmen Sandra Guzmán Calcina, PhD,

Key points

From the paper's conclusion
  • These results indicate that the detector can provide reliable and reproducible data and demonstrate the applicability of active-matrix technology to experimental 3D dosimetry in radiotherapy.

Abstract

A three-dimensional electronic detector array proof-of-principle prototype based on active-matrix technology was developed for radiotherapy dosimetry. The system consists of a 30×30×15 cm 3 PMMA phantom with 2645 embedded detector elements distributed in five active matrices at five depths. Percent depth-dose (PDD) and profile measurements for a 6 MV photon beam from different radiotherapy linacs were acquired and compared to GEANT 4 Monte Carlo simulations. Deviations were < 1% (MC Detector - modelled detector) and < 3% (MC PMMA - PMMA medium). Measurement versus simulation average difference was 0.5% (MC Detector ) and 1.3% (MC PMMA ) within the irradiated field. Reproducibility across measurements was 0.5%. These results indicate that the detector can provide reliable and reproducible data and demonstrate the applicability of active-matrix technology to experimental 3D dosimetry in radiotherapy.