Evaluation of a multifunctional phantom for geometric, image quality, and dosimetric constancy of megavoltage electronic portal imaging devices
Inbum Lee, Minjae Choi, Yoonsuk Huh, Euntaek Yoon, Bo‐Wi Cheon, Na Hye Kwon, Young Jae Jang, Chang Heon Choi,
Key points
From the paper's conclusion- Conclusions MZ Slab supported geometric, image‐quality, radiation‐field, qualitative MLC, and portal‐dosimetry assessments within a single‐phantom QA workflow.
Abstract
Background Routine quality assurance (QA) of linear accelerators and electronic portal imaging devices (EPIDs) often requires separate devices and acquisition procedures for geometric, image‐quality, radiation‐field, multileaf collimator (MLC), and dosimetric checks. A multifunctional phantom could consolidate these assessments within a single QA workflow. Purpose To evaluate the week‐to‐week reproducibility of measurements obtained with MZ Slab, a multifunctional megavoltage EPID phantom, on three linear accelerators with different EPID configurations. Methods Measurements were performed weekly for four weeks on TrueBeam STx, Halcyon, and Harmony Pro. Each session comprised three acquisition sets with phantom removal and repositioning between sets. The phantom provided measurements of geometric accuracy, image quality, material‐response isotropy, and radiation‐field size, together with qualitative MLC light‐field checks. Weekly values were summarized as mean and within‐session standard deviation, and changes were assessed relative to week 1. Field sizes were compared with open‐field EPID images and IC Profiler measurements. Dose‐calibrated portal images acquired on TrueBeam were compared with the week 1 image using mean response change and gamma passing rates. Results Central‐marker offset ranged from 0.24 to 0.78 mm across units, with a maximum change from week 1 of 0.29 mm. Marker‐distance deviations changed by no more than 0.03 mm. The maximum changes in mean relative contrast transfer function and low‐contrast contrast‐to‐noise ratio were 0.011 (2.6%) and 1.81 (5.0%), respectively, relative to the week 1 baseline. Uniformity deviation and normalized noise changed by up to 0.50 and 0.17 percentage points, and material‐specific normalized response by up to 0.014. Field sizes changed by up to 0.56 mm and agreed with open‐field EPID measurements within 0.3 mm. IC Profiler values were 0.2–0.9 mm larger than MZ Slab values. MLC leaf ends remained within the printed ± 1 mm tolerance. At week 4, mean portal‐image response had increased by 0.77%, with a gamma passing rate of 99.65% at 1%/1 mm. Conclusions MZ Slab supported geometric, image‐quality, radiation‐field, qualitative MLC, and portal‐dosimetry assessments within a single‐phantom QA workflow. Measurements on three linacs characterized within‐session reproducibility and week‐to‐week consistency over four weeks. Longer observation and deliberate perturbation testing are required to establish action limits and fault‐detection sensitivity.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Sensitivity in error detection of an electronic portal imaging device for patient-specific QA in head and neck VMAT plans.Radiological physics and technology · 2026
- A deep learning approach for EPID-based patient-specific quality assurance incorporating daily machine-state variations.Frontiers in oncology · 2026
- Retrospective analysis of machine service logs and EPID-based linac QA to investigate the effects of machine maintenance and calibration events on beam characteristics: A multi-institutional study.Journal of applied clinical medical physics · 2026
- API-enabled integration of Pylinac with QATrack+ for automated EPID-based QA on Elekta linear accelerators: a technical note.Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology · 2026
- End to end dosimetric and geometric verification of single isocenter multitarget stereotactic radiosurgery using an anthropomorphic phantom: external audit validation.Radiological physics and technology · 2026
- The suitability of the EPIbeam system for dosimetric verification of radiotherapy treatment plansPolish Journal of Medical Physics and Engineering · 2026