Dosimetric assessment of radiation fields up to 5 × 5 cm 2 and architectural optimization of a diode-based 3D Detector for radiotherapy

Renzo Ocampo, Dante E. Roa, PhD, Enzo Aucca, Renzo Romero, Miguel Risco-Castillo, Carmen Sandra Guzmán Calcina, PhD, William de Souza Santos, PhD, José Rafael Santos Ferreira, MSc,

Key points

From the paper's conclusion
  • The results demonstrate that the new high-density architecture mitigates the spatial undersampling previously observed, ensuring a reliable volumetric dosimetric assessment of small radiation fields.

Abstract

An initial 3D electronic detector array prototype for radiotherapy, consisting of 2645 diodes (8.0 mm pitch), was modelled with Geant4 and MCNP6.2 Monte Carlo codes to simulate irradiations of 2×2, 3×3, 4×4, and 5×5 cm 2 field sizes with a 6 MV photon beam. Comparisons with Treatment Planning Software (TPS) data revealed maximum percent depth dose (PDD) differences of 2.90% for MCNP6.2 and 1.74% for Geant4. However, spatial resolution limitations in the initial prototype caused dosimetric discrepancies in the high-gradient penumbra regions. Consequently, an optimized high-density prototype featuring variable spatial resolution (3.0 mm central pitch, 6.0 mm peripheral pitch) was developed. The results demonstrate that the new high-density architecture mitigates the spatial undersampling previously observed, ensuring a reliable volumetric dosimetric assessment of small radiation fields.