Dosimetric assessment of radiation fields up to 5 × 5 cm 2 and architectural optimization of a diode-based 3D Detector for radiotherapy
Renzo Ocampo, Dante E. Roa, PhD, Enzo Aucca, Renzo Romero, Miguel Risco-Castillo, Carmen Sandra Guzmán Calcina, PhD, William de Souza Santos, PhD, José Rafael Santos Ferreira, MSc,
Key points
From the paper's conclusion- The results demonstrate that the new high-density architecture mitigates the spatial undersampling previously observed, ensuring a reliable volumetric dosimetric assessment of small radiation fields.
Abstract
An initial 3D electronic detector array prototype for radiotherapy, consisting of 2645 diodes (8.0 mm pitch), was modelled with Geant4 and MCNP6.2 Monte Carlo codes to simulate irradiations of 2×2, 3×3, 4×4, and 5×5 cm 2 field sizes with a 6 MV photon beam. Comparisons with Treatment Planning Software (TPS) data revealed maximum percent depth dose (PDD) differences of 2.90% for MCNP6.2 and 1.74% for Geant4. However, spatial resolution limitations in the initial prototype caused dosimetric discrepancies in the high-gradient penumbra regions. Consequently, an optimized high-density prototype featuring variable spatial resolution (3.0 mm central pitch, 6.0 mm peripheral pitch) was developed. The results demonstrate that the new high-density architecture mitigates the spatial undersampling previously observed, ensuring a reliable volumetric dosimetric assessment of small radiation fields.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of a 3D Electronic Detector Array for Radiotherapy Quality AssuranceJournal of Physics: Conference Series · 2026
- A conical-reflector scintillation device for 3-dimensional measurement of radiotherapy beams: a feasibility study.Physical and engineering sciences in medicine · 2026
- Performance evaluation of a new type of commercial silicon detectors for photon beam dosimetryJournal of Applied Clinical Medical Physics · 2026
- Comparison of percentage depth dose measurements obtained with three detector technologies and Monaco TPS calculations in a 1 × 1 cm² small photon field.Biomedical physics & engineering express · 2026
- Toward dynamic Monte Carlo source modeling: measurement variability and primary electron source-width sensitivity in very small photon fields.Frontiers in oncology · 2026
- Scintillating Micro-Probe Detectors for Conventional and FLASH Proton Dosimetry: Optical Transport and Geometry Modeling.Bioengineering (Basel, Switzerland) · 2026