Measuring Small Collimated Proton Fields with a Volumetric Scintillation Detector

Abram Bell, Jason M. Holmes, Sam Beddar, Daniel Robertson

Key points

From the paper's conclusion
  • This system demonstrates potential as a patient quality assurance tool for small, collimated proton fields.

Abstract

Proton stereotactic radiosurgery requires validation of dose distributions and aperture shapes prior to patient treatment. Most two-dimensional (2D) detector arrays lack sufficient spatial resolution to adequately characterize these fields, and only sample the dose distribution at a single depth. In this study, a volumetric scintillator detector is evaluated for validation of small collimated proton fields. 2D projections of three-dimensional (3D) light distributions are measured with a digital camera, and ionization quenching is corrected using Monte Carlo-derived linear energy transfer calculations. The resulting projected dose distributions are compared to dose projections from the treatment planning system. The measured and calculated dose projections agreed with a Gamma passing rate (2%, 2mm) greater than 99%. This system demonstrates potential as a patient quality assurance tool for small, collimated proton fields.