An experimental imaging dose estimation method for the phase-II proton CT scanner.
Lei Chen, Robert P Johnson, Mark Pankuch, Jonathan Bortfeldt, Katia Parodi, Reinhard Schulte, Guillaume Landry, George Dedes
Department of Radiation Oncology, LMU University Hospital, LMU Medizin, LMU Munich, Munich, 81377, Germany. · Department of Physics, UC Santa Cruz, Santa Cruz, CA 95064, United States of America.
Key points
From the paper's conclusion- This study introduces a reliable workflow for imaging dose estimation in the phase-II pCT scanner. The approach demonstrates satisfactory agreement with IC measurements, supporting its feasibility for evaluating imaging dose in image-guided and adaptive proton therapy.
Abstract
PURPOSE: Proton computed tomography (pCT) improves range accuracy in proton therapy by directly mapping proton stopping power relative to water (RSP). Imaging dose remains a crucial factor in determining the feasibility of daily image guidance. This study presents an experimental methodology for estimating imaging dose in the phase-II pCT scanner, utilizing phase space data recorded at the front tracker to eliminate phantom-specific corrections. The estimated doses are validated against ionization chamber (IC) measurements. METHODS: Dose measurements were conducted at multiple locations and angles within two phantoms, with data acquisition triggered by either the front trackers or the energy detector. Baseline scans were also performed without phantoms. Monte Carlo (MC) simulations were used to estimate imaging dose in single projections and full scans. Eight front-tracker-triggered baseline datasets were processed for proton track reconstruction and fluence estimation. After assessing inter-run variability, phase space files were averaged to generate an input source for MC simulations. The simulated imaging doses were renormalized to the averaged datasets and quantitatively compared against IC measurements. RESULTS: Analyzed proton track numbers exhibited consistency across datasets, with coefficients of variation below 2.0%. Estimated doses agreed with IC measurements within ±1.8% for the CTDI phantom; head phantom underestimated -4.6% (single projection) and -6.7% (full scan). Uncertainties from multiple measurements remained below 1.9%. CONCLUSION: This study introduces a reliable workflow for imaging dose estimation in the phase-II pCT scanner. The approach demonstrates satisfactory agreement with IC measurements, supporting its feasibility for evaluating imaging dose in image-guided and adaptive proton therapy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of beam hardening on proton stopping-power ratio estimation using photon-counting, dual-energy, and single-energy computed tomography.Physics and imaging in radiation oncology · 2026
- Patient-Specific Stopping Power Calibration of Treatment Planning X-Ray CT Using Integrated Depth-Dose Profiles Acquired by Proton Transmission Imaging - A Feasibility Study.Physics in medicine and biology · 2026
- Multi-center comparison of relative stopping power estimation of an ex vivo phantom using clinical X-ray CT scanners and a proton CT prototype.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026
- Photon-Counting Computed Tomography Enables Image-Guided Navigation of the Paranasal Sinuses at Ultra-Low Radiation Dose: An Initial Cadaveric Study.Diagnostics (Basel, Switzerland) · 2026
- easyPET.3D Measurement of Positron Emitter Maps in Proton-Irradiated PMMA Phantoms.Biomedical physics & engineering express · 2026
- Towards real-time ion range verification via hybrid Compton-PET 3D imaging at isochronous cyclotronsarXiv · 2026