Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR Brachytherapy
Kevin Willy, Megan Clark, Petr Bruza, Shauna McVorran, David Gladstone
Key points
From the paper's conclusion- We characterize a scintillator-based optical dosimeter integrated into a transrectal probe that tracks source position, speed, dose rate, and time-resolved cumulative dose.
Abstract
High-dose-rate (HDR) brachytherapy is highly conformal, but its accuracy depends on catheter and source placement, and clinical workflows lack real-time in vivo dosimetry (IVD) that verifies delivery without disrupting the procedure. We characterize a scintillator-based optical dosimeter integrated into a transrectal probe that tracks source position, speed, dose rate, and time-resolved cumulative dose. A 67 mm x 5 mm line scintillator is imaged by a CMOS camera through a 30-degree endoscopic lens; geometric and flat-field corrections remove barrel distortion and vignetting. Absolute dose was calibrated with a TOPAS Monte Carlo model of the probe's air-cavity geometry. The system was validated in a water phantom with an Ir-192 source and needles at radial offsets of 7, 10, and 15 mm. Response varied by $\pm$2\% along the 67 mm scintillator and was linear over 5-21 cGy/s ($R^2 \geq 0.995$). The Monte Carlo model showed a 3.75\% dose-rate reduction relative to TG-43 water-equivalent conditions. In phantom, the system resolved a 3 mm needle displacement, confirmed by CT. Measured dose-rate profiles agreed with Monte Carlo in peak magnitude and spatial distribution, and integral dose agreed within 1\%. By mapping dose directly to the measurement plane of a standard ultrasound transducer, this system provides high-resolution, time-resolved verification of HDR delivery and a clinical tool for detecting delivery errors with minimal changes to existing treatment protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Preliminary feasibility results for implementing a phantom model for external auditing of HDR brachytherapy with an Ir-192 HDR source: an IAEA collaborative project.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Distance-dependent influence of scoring voxel size on Monte Carlo dosimetry of an iridium-192 HDR brachytherapy sourceEPJ Web of Conferences · 2026
- A scintillator-based approach to spatial and temporal dosimetry in UHDR electron irradiationJournal of Physics: Conference Series · 2026
- Validation of a detector-response correction model for in vivo dosimetry in high-dose-rate prostate brachytherapy.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- In vivo Dose Measurement in Cervical Cancer Brachytherapy Using Dysprosium-doped Calcium Sulfate Thermoluminescent Dosimeters: A Phantom Study.Journal of medical physics · 2026
- A conical-reflector scintillation device for 3-dimensional measurement of radiotherapy beams: a feasibility study.Physical and engineering sciences in medicine · 2026