Distance-dependent influence of scoring voxel size on Monte Carlo dosimetry of an iridium-192 HDR brachytherapy source

Yacine El Attaoui, Aziz Oustous, Yassine Oustous, Lhoucine Oufni

Key points

From the paper's conclusion
  • We examined how the size of the scoring voxel changes the TG-43 parameters of the Flexisource iridium-192 HDR source, and whether this size should vary with distance.

Abstract

In Monte Carlo dosimetry of high-dose-rate (HDR) brachytherapy, the scoring grid has to be small enough to follow the sharp dose fall-off near the source, but large enough for the statistical noise to stay low at large distances. We examined how the size of the scoring voxel changes the TG-43 parameters of the Flexisource iridium-192 HDR source, and whether this size should vary with distance. The source was simulated in TOPAS v3.9 (Geant4 10.07); g L ( r ) and F ( r ,θ) were tallied in water with cubic voxels of 0.5 mm and 1.0 mm side and compared with the CLRP v2 data. We obtained Λ = 1.098 ± 0.025 cGy·h −1 ·U −1 , 1.3% below the consensus value. Within 3 cm of the source the 0.5 mm grid reproduced g L ( r ) with a mean absolute deviation of 0.50% (maximum 1.78%), while the 1 mm grid was off by as much as 20.0%; further out the order was inverted (2.58% against 0.56% on average). Switching from the fine to the coarse grid at about 3 cm limited the deviation to 1.78% at all distances.