Toward dynamic Monte Carlo source modeling: measurement variability and primary electron source-width sensitivity in very small photon fields.
Mina M Habib
Department of Radiation Physics, Maadi Armed Forces Medical Compound, Cairo, Egypt.
Key points
From the paper's conclusion- The M1 PDD benchmark supported reference depth-dose behaviour, whereas lateral-profile findings were restricted to controlled numerical sensitivity because absolute lateral agreement was not independently established in this study alone.
Abstract
BACKGROUND: Very small photon fields are sensitive to both measurement conditions and radiation-source modelling, but the connection between effective primary-electron source width, downstream photon angular phase space, and field-size-dependent dosimetric response remains incompletely characterized. METHODS: The study was restricted to 6 MV. Empirical measurement variability was evaluated using five ionization-chamber datasets acquired on a clinical Elekta Versa HD for fields from 20×20 to 3 × 3 cm2. A generalized Versa HD/Agility EGSnrc/BEAMnrc model was evaluated with two prescribed stationary Gaussian source endpoints, M1 (FWHM x = FWHM y = 0.15 cm) and M2 (FWHM x = 0.35 cm, FWHM y = 0.60 cm). DOSXYZnrc dose responses and final pre-phantom photon phase-space data were analysed for 3×3, 2×2, 1×1, and 0.5×0.5 cm2. The 3×3 cm2 field provided the experimental-Monte Carlo overlap; the smaller fields were controlled Monte Carlo endpoint-sensitivity analyses. M1 was additionally benchmarked against a 10×10 cm2 CC13 commissioning PDD. RESULTS: Mean experimental coefficient of variation increased from 2.46% at 20×20 cm2 to 15.45% at 3×3 cm2. The M2 central-axis reductions were 8.48%, 8.27%, 29.59%, and 56.34% for 3×3, 2×2, 1×1, and 0.5×0.5 cm2, respectively. The corresponding M2/M1 ratios of the 95% angular-containment solid angle were 1.275, 1.498, 2.415, and 3.835. The four-field comparison demonstrates clear field-size progression in angular containment, while showing that amplification is observable-dependent rather than universally monotonic. The M1 PDD benchmark supported reference depth-dose behaviour, whereas lateral-profile findings were restricted to controlled numerical sensitivity because absolute lateral agreement was not independently established in this study alone. CONCLUSION: The demonstrated result is a stationary endpoint-sensitivity chain linking effective primary-electron source representation, linac-head transport, photon angular phase space, and small-field dose/profile response. A reduced covariance formulation is retained only as a prospective framework for testing whether a stationary effective source can represent unresolved accelerator-source states; no time-resolved source dynamics were measured or simulated.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of percentage depth dose measurements obtained with three detector technologies and Monaco TPS calculations in a 1 × 1 cm² small photon field.Biomedical physics & engineering express · 2026
- Performance evaluation of a new type of commercial silicon detectors for photon beam dosimetryJournal of Applied Clinical Medical Physics · 2026
- Dosimetric comparison of 6-MV flattened and flattening filter-free photon beams using Monte Carlo simulationNuclear Engineering and Technology · 2026
- Monte Carlo evaluation of small-field output correction factors of a microDiamond detector in flattened and flattening-filter-free photon beams for stereotactic radiotherapy.Radiation and environmental biophysics · 2026
- Distance-dependent influence of scoring voxel size on Monte Carlo dosimetry of an iridium-192 HDR brachytherapy sourceEPJ Web of Conferences · 2026
- Pulse width-dependent Monte Carlo source modelling for ultra-high dose rate electron beams.Physics in medicine and biology · 2026