EPJ Web of Conferences

Experimental and Monte Carlo TG-43 characterization of the HDR 192 Ir Flexisource: A combined EBT3 film and GATE study

Aziz Oustous, Zouhair Airouss, Yacine El Attaoui, Yassine Oustous, Lhoucine Oufni

Key points

From the paper's conclusion
  • Both methods gave Λ = 1.078 cGy h −1 U −1 , 2.79% lower than Granero et al. g L ( r ) agreed with the reference within 3.43% (GATE) and 2.3% (film), and F ( r ,θ) within 3% and 2.9% for 15° < θ < 175°; larger differences appeared near the long axis, for different reasons in each method.

Abstract

TG-43U1 asks that the dataset of a brachytherapy source be based on both a Monte Carlo calculation and a measurement. For one HDR 192 Ir Flexisource, we compared a GATE/Geant4 simulation with Gafchromic EBT3 film dosimetry, and checked how well the two independent methods agree with each other and with the literature. The source was modelled in GATE v9.1 inside a water sphere of 20 cm radius; films from one batch were exposed in a 30×30×30 cm 3 PMMA phantom with full backscatter. Both methods gave Λ = 1.078 cGy h −1 U −1 , 2.79% lower than Granero et al. g L ( r ) agreed with the reference within 3.43% (GATE) and 2.3% (film), and F ( r ,θ) within 3% and 2.9% for 15° < θ < 175°; larger differences appeared near the long axis, for different reasons in each method. The agreement of two independent techniques supports both the GATE model and the film protocol for this source.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers