Experimental and Monte Carlo TG-43 characterization of the HDR 192 Ir Flexisource: A combined EBT3 film and GATE study
Aziz Oustous, Zouhair Airouss, Yacine El Attaoui, Yassine Oustous, Lhoucine Oufni
Key points
From the paper's conclusion- Both methods gave Λ = 1.078 cGy h −1 U −1 , 2.79% lower than Granero et al. g L ( r ) agreed with the reference within 3.43% (GATE) and 2.3% (film), and F ( r ,θ) within 3% and 2.9% for 15° < θ < 175°; larger differences appeared near the long axis, for different reasons in each method.
Abstract
TG-43U1 asks that the dataset of a brachytherapy source be based on both a Monte Carlo calculation and a measurement. For one HDR 192 Ir Flexisource, we compared a GATE/Geant4 simulation with Gafchromic EBT3 film dosimetry, and checked how well the two independent methods agree with each other and with the literature. The source was modelled in GATE v9.1 inside a water sphere of 20 cm radius; films from one batch were exposed in a 30×30×30 cm 3 PMMA phantom with full backscatter. Both methods gave Λ = 1.078 cGy h −1 U −1 , 2.79% lower than Granero et al. g L ( r ) agreed with the reference within 3.43% (GATE) and 2.3% (film), and F ( r ,θ) within 3% and 2.9% for 15° < θ < 175°; larger differences appeared near the long axis, for different reasons in each method. The agreement of two independent techniques supports both the GATE model and the film protocol for this source.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Distance-dependent influence of scoring voxel size on Monte Carlo dosimetry of an iridium-192 HDR brachytherapy sourceEPJ Web of Conferences · 2026
- Clinical test cases for commissioning, QA, and benchmarking of model‐based dose calculation algorithms in¹ ⁹² Ir HDR gynecologic tandem and ring brachytherapyMedical Physics · 2026
- Applicator-dependent differences between TG-43 and a grid-based Boltzmann solver in iridium-192 high-dose-rate gynecological brachytherapy.Journal of contemporary brachytherapy · 2026
- Investigation of TG-43 dose calculation algorithm for volumetric HDR Brachytherapy using Monte Carlo simulation in TOPAS.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Monte Carlo intercomparison of TG-43 dosimetric parameters for clinical192Ir and125I brachytherapy sources.Physics in medicine and biology · 2026
- Dose distribution in a cervix phantom quality-assurance using two computer modeling approaches.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026