Journal of contemporary brachytherapy

Applicator-dependent differences between TG-43 and a grid-based Boltzmann solver in iridium-192 high-dose-rate gynecological brachytherapy.

Bojan Pavičar, Andrej Milosavljević, Branko Predojević, Borislava Petrović

Faculty of Sciences, Department of Physics, University of Novi Sad, 21000 Novi Sad, Serbia. · Radiotherapy Department, International Medical Centers, 78000 Banja Luka, Bosnia and Herzegovina.

Observational · n=40

Key points

From the paper's conclusion
  • TG-43 systematically overestimated dose relative to Acuros® BV in all applicator groups. Differences were smallest for T&R, largest and most variable for UMC, and intermediate for T&O and SC. Algorithm-related dose differences are applicator-dependent and should not be assumed uniform across treatment configurations.

Abstract

PURPOSE: To compare TG-43 and Acuros® BV dose estimates in iridium-192 (192Ir) high-dose-rate (HDR) gynecological brachytherapy across four applicator types and evaluate applicator-dependent differences. The novelty was the systematic comparison of multiple applicators within a single-institution workflow. MATERIAL AND METHODS: This retrospective dosimetric study included 40 patients and 121 treatment plans: tandem and ring (T&R, 34 plans), tandem and ovoid (T&O, 38 plans), segmented cylinder (SC, 23 plans), and universal multi-channel cylinder (UMC, 26 plans). All plans were generated in BrachyVision™ v. 15.6 using the TG-43 formalism and subsequently recalculated with Acuros® BV, with identical dwell positions and dwell times. The main analysis included high-risk clinical target volume (HR-CTV) D90, D98, and V100; organ-at-risk (OAR) D2cc; Point A for T&R/T&O; and the SC 100% prescription-isodose position at 5 mm depth. Exploratory dwell-loading descriptors and a density-cap quality assurance audit supported physical and methodological interpretation. RESULTS: TG-43 yielded consistently higher dose estimates than Acuros® BV. For HR-CTV D90, relative differences were 2.9 ±0.5% for T&R, 3.7 ±0.8% for T&O, 3.2 ±0.4% for SC, and 4.9 ±2.2% for UMC. OAR D2cc differences were also largest for UMC, ranging from 4.3% (3.4-7.1%) in bladder to 6.1 ±2.2% in small bowel. For SC, when the TG-43 100% prescription isodose was normalized to 5 mm from the applicator surface, the corresponding Acuros® BV 100% isodose position was 4.6 ± 0.1 mm, indicating an inward shift of 0.4 ±0.1 mm. CONCLUSIONS: TG-43 systematically overestimated dose relative to Acuros® BV in all applicator groups. Differences were smallest for T&R, largest and most variable for UMC, and intermediate for T&O and SC. Algorithm-related dose differences are applicator-dependent and should not be assumed uniform across treatment configurations.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers