Applicator-dependent differences between TG-43 and a grid-based Boltzmann solver in iridium-192 high-dose-rate gynecological brachytherapy.
Bojan Pavičar, Andrej Milosavljević, Branko Predojević, Borislava Petrović
Faculty of Sciences, Department of Physics, University of Novi Sad, 21000 Novi Sad, Serbia. · Radiotherapy Department, International Medical Centers, 78000 Banja Luka, Bosnia and Herzegovina.
Key points
From the paper's conclusion- TG-43 systematically overestimated dose relative to Acuros® BV in all applicator groups. Differences were smallest for T&R, largest and most variable for UMC, and intermediate for T&O and SC. Algorithm-related dose differences are applicator-dependent and should not be assumed uniform across treatment configurations.
Abstract
PURPOSE: To compare TG-43 and Acuros® BV dose estimates in iridium-192 (192Ir) high-dose-rate (HDR) gynecological brachytherapy across four applicator types and evaluate applicator-dependent differences. The novelty was the systematic comparison of multiple applicators within a single-institution workflow. MATERIAL AND METHODS: This retrospective dosimetric study included 40 patients and 121 treatment plans: tandem and ring (T&R, 34 plans), tandem and ovoid (T&O, 38 plans), segmented cylinder (SC, 23 plans), and universal multi-channel cylinder (UMC, 26 plans). All plans were generated in BrachyVision™ v. 15.6 using the TG-43 formalism and subsequently recalculated with Acuros® BV, with identical dwell positions and dwell times. The main analysis included high-risk clinical target volume (HR-CTV) D90, D98, and V100; organ-at-risk (OAR) D2cc; Point A for T&R/T&O; and the SC 100% prescription-isodose position at 5 mm depth. Exploratory dwell-loading descriptors and a density-cap quality assurance audit supported physical and methodological interpretation. RESULTS: TG-43 yielded consistently higher dose estimates than Acuros® BV. For HR-CTV D90, relative differences were 2.9 ±0.5% for T&R, 3.7 ±0.8% for T&O, 3.2 ±0.4% for SC, and 4.9 ±2.2% for UMC. OAR D2cc differences were also largest for UMC, ranging from 4.3% (3.4-7.1%) in bladder to 6.1 ±2.2% in small bowel. For SC, when the TG-43 100% prescription isodose was normalized to 5 mm from the applicator surface, the corresponding Acuros® BV 100% isodose position was 4.6 ± 0.1 mm, indicating an inward shift of 0.4 ±0.1 mm. CONCLUSIONS: TG-43 systematically overestimated dose relative to Acuros® BV in all applicator groups. Differences were smallest for T&R, largest and most variable for UMC, and intermediate for T&O and SC. Algorithm-related dose differences are applicator-dependent and should not be assumed uniform across treatment configurations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Investigation of TG-43 dose calculation algorithm for volumetric HDR Brachytherapy using Monte Carlo simulation in TOPAS.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Dosimetric comparison of HDR Co-60 and Ir-192 brachytherapy sources by model-based dose calculation algorithms and its clinical implication in different brachytherapy sites and techniques.Brachytherapy · 2026
- Clinical test cases for commissioning, QA, and benchmarking of model‐based dose calculation algorithms in¹ ⁹² Ir HDR gynecologic tandem and ring brachytherapyMedical Physics · 2026
- Experimental and Monte Carlo TG-43 characterization of the HDR 192 Ir Flexisource: A combined EBT3 film and GATE studyEPJ Web of Conferences · 2026
- Distance-dependent influence of scoring voxel size on Monte Carlo dosimetry of an iridium-192 HDR brachytherapy sourceEPJ Web of Conferences · 2026
- A Monte Carlo based dose verification tool for an HDR Ir-192 brachytherapy source and its evaluation on extremity malignancies.Journal of applied clinical medical physics · 2026