Capturing human-machine interplay in automated insulin delivery systems through day-to-day variability of the automated correction index (ACI-V%).
Andrea Tumminia, Giuseppe Papa, Rossella Cannarella, Francesco Galeano, Vittorio Oteri, Vincenzo Amato, Roberto Baratta, Rosita A Condorelli,
Endocrine Unit, Garibaldi-Nesima Hospital, Catania, Italy. Electronic address: andreatumminia82@gmail.com. · Unit of Metabolic and Endocrine Disease, "Centro Catanese di Medicina e Chirurgia" Clinic, Catania, Italy. Electronic address: gpapa_98@yahoo.com.
Key points
From the paper's conclusion- ACI-V% provides an exploratory descriptor of day-to-day variability in automated correction insulin delivery that may complement mean ACI for clinical interpretation of AHCL system behavior.
Abstract
BACKGROUND: The Automated Correction Index (ACI) reflects the proportion of algorithm-driven correction boluses relative to total bolus insulin in advanced hybrid closed-loop (AHCL) systems. While mean ACI has been associated with glycemic outcomes in patients with type 1 diabetes mellitus (T1DM), the clinical significance of its day-to-day variability (ACI-V%) remains unknown. METHODS: In this multicenter real-world study, 330 pediatric and adult patients with T1DM using the MiniMed™ 780G system were studied. ACI-V% was defined as the coefficient of variation of daily ACI values over a 30-day period. Associations between ACI-V% and continuous glucose monitoring (CGM) metrics were assessed. RESULTS: Higher ACI-V% was associated with greater time in range (TIR70-180 mg/dL) and lower time above range, while no significant association with time below range was observed. ACI-V% showed an inverse correlation with mean ACI, suggesting greater day-to-day variability in settings of lower overall automated correction insulin delivery. Moreover, ACI-V% independently predicted achievement of TIR70-180 mg/dL > 70%, with a threshold of ≥ 32.6% identifying patients more likely to reach this target (sensitivity 52.4%, specificity 71.4%). CONCLUSIONS: ACI-V% provides an exploratory descriptor of day-to-day variability in automated correction insulin delivery that may complement mean ACI for clinical interpretation of AHCL system behavior.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The Varying Benefits of Two Years Automated Insulin Delivery System Use on Continuous Glucose Monitoring Profiles in Children and Adolescents with Type 1 Diabetes.Diabetes technology & therapeutics · 2026
- The Automatic Correction Boluses in an Advanced Hybrid Closed-Loop System: Algorithm Performance and Clinical ImplicationsJournal of Diabetes Science and Technology · 2026
- Insulin delivery characteristics and glycemic control in older adults with type 1 diabetes using the Minimed™ 780G system: a real-world analysis.Frontiers in endocrinology · 2026
- Glycemic Outcomes of Twiist™ Automated Insulin Delivery System in People with Diabetes Requiring Insulin, A Large Real-World Study.Diabetes technology & therapeutics · 2026
- Narrowing the A1c gap: Personalized modeling of HbA1c– continuous glucose monitor discordance in type 1 diabetesPLOS Digital Health · 2026
- Estimated glucose disposal rate is associated with continuous glucose monitoring metrics in type 1 diabetesScientific Reports · 2026