Diabetes technology & therapeutics

The Varying Benefits of Two Years Automated Insulin Delivery System Use on Continuous Glucose Monitoring Profiles in Children and Adolescents with Type 1 Diabetes.

Jantje Weiskorn, Marie Auzanneau, Marie-Anne Burckhardt, Nicole Blauensteiner, Ulrike Schierloh, Isa Mayer, Julia Schneidewind, Stefanie Lanzinger

Auf der Bult, Diabetes Center for Children and Adolescents, Hannover, Germany. · Institute of Epidemiology and Medical Biometry, Ulm University, Ulm, Germany.

Observational · n=2,486

Key points

From the paper's conclusion
  • After 2 years of AID use, TiTR and CV showed clinically relevant improvements, while TBR < 70 mg/dL increased with stable TBR < 54 mg/dL and severe hypoglycemia decreased. Changes in TIR, SG, and HbA1c were statistically significant but small, yet noteworthy, as glycemia usually deteriorates during adolescence. The increase in BMI-SDS needs to be further investigated.

Abstract

BACKGROUND: Automated insulin delivery (AID) systems are recommended for youths with type 1 diabetes, but real-world longitudinal data remains limited despite increasing use. This study evaluates the effect of 2 years of AID use on continuous glucose monitoring (CGM) profiles and body mass index (BMI) development. METHODS: Data from individuals with type 1 diabetes aged <18 years, diabetes duration >3 months, and AID use of >3 months from the Diabetes Prospective Follow-up (DPV) Registry between 2019 and 2025 were analyzed. CGM and HbA1c were required at the last documented time in range (TIR) (±90 days; follow-up) and 2 years earlier (±90 days; baseline). Changes were investigated using adjusted regression models with repeated measurements. FINDINGS: Data from 2486 participants (baseline: median age: 11.9 years, diabetes duration: 5.0 years, 53% male) were analyzed over ∼23 months. The largest gains occurred in time in tight range (TiTR) (42.4% [95% confidence interval: 41.0-43.7] to 50.1% [48.8-51.4]; P < 0.001) and coefficient of variation (CV) (32.9 [32.0-33.7] to 29.5 [28.7-30.2]; P < 0.001). HbA1c (7.3% [7.2-7.3] to 7.2 [7.1-7.2]; P < 0.001), TIR (64.5% [63.9-65.1] to 66.7% [66.1-67.3]; P < 0.001), and mean sensor glucose (SG) (161.0 mg/dL [158.4-163.5] to 155.5 mg/dL [153.1-157.9]; P = 0.001) improved statistically, but below clinical relevance thresholds. Time below range (TBR) < 70 mg/dL increased (2.7% [2.6-2.9] to 3.2% [3.1-3.4]; P < 0.001), TBR < 54 mg/dL remained stable (1.7%-1.8%; P = 0.516), and severe hypoglycemia decreased (6.4-4.5/100 patient-years; P < 0.001). No changes were seen in children <6 years. In youth 6 to <12 and 12 to <18 years, TiTR and CV improved most, while TBR increased. BMI-standard deviation score (SDS) increased (0.57 [0.53-0.61] to 0.67 [0.63-0.72]; P < 0.001). CONCLUSION: After 2 years of AID use, TiTR and CV showed clinically relevant improvements, while TBR < 70 mg/dL increased with stable TBR < 54 mg/dL and severe hypoglycemia decreased. Changes in TIR, SG, and HbA1c were statistically significant but small, yet noteworthy, as glycemia usually deteriorates during adolescence. The increase in BMI-SDS needs to be further investigated.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers