Glycemic Outcomes of Twiist™ Automated Insulin Delivery System in People with Diabetes Requiring Insulin, A Large Real-World Study.
Viral N Shah, Peter G Jacobs, Grazia Aleppo, Anders Carlson, Carol J Levy, Ken Snow, Joanna Mitri
Division of Endocrinology & Metabolism, Indiana University School of Medicine, Indianapolis, Indiana, USA. · Oregon Health & Science University, Portland, Oregon, USA.
Key points
From the paper's conclusion- In this large real-world cohort, use of the twiist AID system was associated with effective overall and overnight glycemic management and minimal hypoglycemia.
Abstract
BACKGROUND: Automated insulin delivery (AID) systems are recommended for individuals with diabetes requiring insulin. Real-world effectiveness may differ from outcomes reported in prospective clinical trials. The twiist™ AID system recently received regulatory clearance, and real-world glycemic outcomes have not been previously described. METHODS: This was an ongoing retrospective, observational real-world study with cross-sectional analyses of participant-level summary data. The analysis included individuals with type 1 diabetes (T1D) and type 2 diabetes (T2D) who initiated use of the twiist AID system and consented to data sharing. Deidentified continuous glucose monitoring (CGM) and insulin delivery data were analyzed. Inclusion in the analysis required ≥14 days of device use. Glycemic outcomes were assessed using standard CGM metrics, including percent time in range (%TIR; 70-180 mg/dL), time in tight range (%TITR; 70-140 mg/dL), time below range (%TBR; 54 to <70 and <54 mg/dL), and time above range (%TAR; >180-250 and >250 mg/dL). Subgroup analyses evaluated associations with baseline glycated hemoglobin (HbA1c), age, and bolus frequency. RESULTS: Data from 4000 individuals were included. The duration of device use ranged from a minimum of 14 days to a maximum of 365 days. Mean overall %TIR was 73.5 ± 14.5%, with overnight %TIR of 77.9 ± 16.5%. Hypoglycemia exposure was low, with mean overall %TBR of 54 to <70 mg/dL of 1.9 ± 2.0% and <54 mg/dL of 0.2 ± 0.4%. Glycemic outcomes were generally within recommended targets across age categories except for %TIR among those <18 years of age. Among those with baseline HbA1c values and TIR values (n = 1131), %TIR was 81.5 ± 10.2% and 71.1 ± 11.4%, and 61.7 ± 16.3% for those HbA1c <7%, 7%-8%, and >8%, respectively. More frequent boluses were associated with improved glycemic outcomes. CONCLUSIONS: In this large real-world cohort, use of the twiist AID system was associated with effective overall and overnight glycemic management and minimal hypoglycemia.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Multicenter 13-Week Evaluation of the twiist Automated Insulin Delivery in Insulin-Treated Type 2 Diabetes: The twiist-T2D Pivotal Trial.Diabetes care · 2026
- Sustained Improvement in Time in Range After Automated Insulin Delivery Initiation in Adolescents and Adults With Type 1 Diabetes Mellitus: A 12-Month Prospective Observational Study From Saudi Arabia.Journal of diabetes science and technology · 2026
- The Varying Benefits of Two Years Automated Insulin Delivery System Use on Continuous Glucose Monitoring Profiles in Children and Adolescents with Type 1 Diabetes.Diabetes technology & therapeutics · 2026
- Real-World Outcomes of Automated Insulin Delivery in Young People With Type 1 Diabetes: A Systematic Review and Meta-Analysis.Diabetes, obesity & metabolism · 2026
- A nationwide 12‐month observatory of automated insulin delivery shows improved glucose control, sustained adoption, and reduced acute severe eventsDiabetes Obesity and Metabolism · 2025
- Insulin delivery characteristics and glycemic control in older adults with type 1 diabetes using the Minimed™ 780G system: a real-world analysis.Frontiers in endocrinology · 2026