Insulin delivery characteristics and glycemic control in older adults with type 1 diabetes using the Minimed™ 780G system: a real-world analysis.
Špela Volčanšek, Petra Vidmar, Andrej Janež, Aleš Skvarča
University Medical Centre Ljubljana, Department of Endocrinology, Diabetes and Metabolic Diseases, Ljubljana, Slovenia.
Key points
From the paper's conclusion- While the MiniMed™ 780G AID system use is associated with low CGM-derived hypoglycemia exposure and highly effective glycemic control, regardless of customized settings, superior glycemic control is associated with a higher proportion of user-initiated manual boluses. The algorithm supports, but cannot replace, user engagement.
Abstract
INTRODUCTION: Diabetes self-management in older adults with type 1 diabetes (T1D) presents a unique clinical challenge, since maintaining tight glycemic control must be carefully weighed against age-related vulnerabilities. This study aimed to analyze glycemic control through the effects of customized algorithm settings in a cohort of older adults with T1D using an automated insulin delivery (AID) system. METHODS: This retrospective real-world data analysis included users of the MiniMed™ 780G AID system, aged ≥60 years. Continuous glucose monitoring (CGM) data including time in range (TIR; 70-180 mg/dL; 3.9-10.0 mmol/L), time in tight range (TITR; 70-140 mg/dL; 3.9-7.8 mmol/L), time below range (TBR; <70 mg/dL; <3.9 mmol/L), level 2 TBR (<54 mg/dL; <3.0 mmol/L) and time above range (TAR; >180 mg/dL; >10.0 mmol/L) were analyzed alongside algorithm settings and daily insulin use. Group comparisons utilized standard parametric and non-parametric statistical tests. A multivariable logistic regression model identified independent clinical predictors of high glycemic performance (TITR ≥50%). RESULTS: The cohort comprised 97 AID users (70.1% female, mean age 67.9 ± 6.7; range 60-90 years). Overall glycemic control was good, with mean TIR 76.1%, TITR 51.1% and TBR 1.1%. Fourteen participants (14.4%) used the manufacturer's recommended optimal settings (ROS); i.e., target glucose of 5.5 mmol/L and active insulin time of 2 hours. CGM metrics did not differ between ROS and non-ROS users for TIR, TITR and TBR. When stratifying the cohort by glycemic performance, 55 participants (56.7%) reached TITR ≥50%; however, the adoption of ROS in high (≥50%) TITR and lower (<50%) TITR subgroups was similar (16.4% vs 11.9%; Fisher p=0.58). When compared to the lower TITR subgroup, participants in the high TITR subgroup delivered less auto-correction insulin (14.2% vs 20.6%, p<0.001) and a higher proportion of manual bolus insulin (46.1% vs 35.2%, p<0.001), which remained a significant predictor of high TITR after multivariable adjustment. TBR was significantly higher in the high (≥50%) TITR subgroup, although it remained low in absolute terms (1.5% vs. 0.8%; p=0.017). CONCLUSION: While the MiniMed™ 780G AID system use is associated with low CGM-derived hypoglycemia exposure and highly effective glycemic control, regardless of customized settings, superior glycemic control is associated with a higher proportion of user-initiated manual boluses. The algorithm supports, but cannot replace, user engagement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Contemporary automated insulin delivery systems in older adults with type 1 diabetes: A systematic review and meta-analysis.Diabetes research and clinical practice · 2026
- Glycemic and psychosocial outcomes of automated insulin delivery in older and high-risk populations with type 1 diabetes: a systematic review and meta-analysis.The Journal of clinical endocrinology and metabolism · 2026
- Automated insulin delivery in older adults with type 1 diabetes: evidence, limitations, and clinical perspectives.Endocrine · 2026
- Glycemic Outcomes of Twiist™ Automated Insulin Delivery System in People with Diabetes Requiring Insulin, A Large Real-World Study.Diabetes technology & therapeutics · 2026
- Sustained Improvement in Time in Range After Automated Insulin Delivery Initiation in Adolescents and Adults With Type 1 Diabetes Mellitus: A 12-Month Prospective Observational Study From Saudi Arabia.Journal of diabetes science and technology · 2026
- Real-world Glycemic Outcomes of a Tubeless Automated Insulin Delivery System (Medtrum A8 TouchCare) in Adults With Type 1 DiabetesCanadian Journal of Diabetes · 2026