The Automatic Correction Boluses in an Advanced Hybrid Closed-Loop System: Algorithm Performance and Clinical Implications
Giuseppe Papa, Rossella Cannarella, Concetta Finocchiaro, Massimiliano Anzaldi, Carmelo Gusmano, Roberto Palazzolo, Gianfranco Gruttadauria, Luca Patti,
University of Catania · Ospedale Cannizzaro
Key points
From the paper's conclusion- Automatic correction boluses (ABs) are a key feature of advanced hybrid closed-loop systems, yet their clinical relevance remains incompletely characterized.
Abstract
BACKGROUND: Automatic correction boluses (ABs) are a key feature of advanced hybrid closed-loop systems, yet their clinical relevance remains incompletely characterized. METHODS: We retrospectively analyzed insulin delivery and continuous glucose monitoring (CGM) data from 287 individuals with type 1 diabetes using the MiniMed 780G for ≥6 months with ≥80% auto-mode usage, recruited from four diabetes centers in Sicily. We quantified the contribution of ABs to total daily insulin dose (TDD) and total bolus insulin and evaluated associations with demographic factors, pump settings, and dietary habits using multivariate regression and receiver operating characteristic (ROC) analyses. RESULTS: ABs accounted for 18.3% of TDD and 33.1% of total bolus insulin. AB/TDD was higher in pediatric participants and inversely associated with diabetes duration and AID use. Active insulin time and carbohydrate intake independently predicted AB/TDD. Higher AB/TDD was associated with lower time in range, higher glucose variability, time above range, and GMI; in pediatric participants, it was also associated with reduced time below range. An AB/TDD ≤20.3% predicted TIR >70% (AUC 0.785) and ≤18.6% predicted GMI <7% (AUC 0.828). CONCLUSIONS: ABs represent a substantial component of insulin delivery in MiniMed 780G users. Higher AB contribution is associated with poorer glycemic outcomes. Identified thresholds may support system optimization in clinical practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- MiniMed 780G Advanced Hybrid Closed-Loop System Use During Days Without User-Initiated BolusesDiabetes Care · 2026
- Sustained Glycemic Outcomes in Children: A Three-Year Real-World Follow-Up of the MiniMed™ 780G System.Hormone research in paediatrics · 2026
- Capturing human-machine interplay in automated insulin delivery systems through day-to-day variability of the automated correction index (ACI-V%).Diabetes research and clinical practice · 2026
- Glycemic And Patient-Reported Outcomes After Transition To Advanced Hybrid Closed-Loop System In Patients After Total Pancreatectomy.The Journal of clinical endocrinology and metabolism · 2026
- Improvement in Tight Glycemic Control in Adults with Type 1 Diabetes Using the MiniMed™ 780G Hybrid Closed-Loop System: A Real-World Evaluation.Current diabetes reviews · 2026
- Prospective multicenter evaluation of glycemic and patient-reported outcomes following transition to a next-generation continuous glucose monitoring system in users of advanced hybrid closed-loop technologyTherapeutic Advances in Endocrinology and Metabolism · 2026