Monte Carlo-based correction of scattering effects in low-energy mass attenuation experimental measurements of 3D-Printed polymers.
João Pedro Carvalho Andrade, William Luna Salgado, César Marques Salgado, Roos Sophia de Freitas Dam
Instituto de Engenharia Nuclear (IEN), Serviço de Desenvolvimento de Técnicas Nucleares (SEDTN), Rio de Janeiro, RJ, Brazil; Universidade Federal do Rio de Janeiro (UFRJ), Escola Politécnica (POLI), Rio de Janeiro, RJ, Brazil. Electronic address: joaoandrade.20221@poli.ufrj.br. · Instituto de Engenharia Nuclear (IEN), Serviço de Desenvolvimento de Técnicas Nucleares (SEDTN), Rio de Janeiro, RJ, Brazil.
Key points
From the paper's conclusion- In this work, a Monte Carlo-based correction methodology using the MCNP6 code was developed to improve the accuracy of mass attenuation coefficient measurements in 3D-printed polymeric materials.
Abstract
Low-energy gamma-ray attenuation measurements are strongly influenced by coherent (Rayleigh) scattering and geometrical effects, particularly under non-ideal broad-beam experimental conditions. In this work, a Monte Carlo-based correction methodology using the MCNP6 code was developed to improve the accuracy of mass attenuation coefficient measurements in 3D-printed polymeric materials. Experimental measurements were performed for acrylonitrile butadiene styrene (ABS), polyethylene terephthalate glycol (PETG), and polylactic acid (PLA) using 241Am and 137Cs gamma-ray sources in combination with a 2 × 2″ NaI(Tl) scintillation detector. Monte Carlo simulations were conducted to reproduce both the experimental measurement geometry and an idealized narrow-beam configuration, enabling the determination of correction factors for low-energy attenuation measurements affected by scattering effects. Additional simulations were performed within the 40-100 keV energy range to develop mathematical fitting functions capable of estimating scattering and geometrical corrections as a function of photon energy. Aluminum samples were also analyzed as a reference material to validate both the experimental methodology and the MCNP6 simulations. The obtained attenuation coefficients were compared with NIST XCOM reference data, showing maximum deviations of 4.94% for the 137Cs measurements (aluminum) and 9.50% for the corrected 241Am measurements (PLA). Results demonstrated that the proposed correction methodology significantly improved agreement between experimental and reference attenuation coefficients, particularly for low-energy photons strongly affected by Rayleigh scattering and geometrical effects. The proposed approach enables reliable attenuation measurements to be performed even under non-ideal experimental conditions, particularly in laboratories without access to highly collimated measurement systems or high-intensity radioactive sources. This methodology contributes to the assessment of 3D-printed polymers as alternative materials for radiation shielding and nuclear technology applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- ABS and PETG as candidate materials for 3D-printed mammography phantoms: Physical and radiological characterization and assessment of TLD-indicated air kerma as a function of depth.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Deformation-Aware Monte Carlo Assessment of a Bi2O3/Thermoplastic Polyurethane Shielding Surrogate Under Tensile Strain and Local Thickness Loss.Polymers · 2026
- Experimental determination of effective X-ray attenuation coefficients of 3D-printed materials under clinical mammography spectraFrontiers in Bioengineering and Biotechnology · 2026
- A multi-thickness regression framework for determining photon attenuation parameters: Application to Sm2O3-doped tellurite glasses using MCNP6.2.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Angle-dependent X-ray absorption correction in small- and wide-angle X-ray scattering: accounting for constrained scattering volume, beam profiles and capillary wall attenuation.Journal of applied crystallography · 2026
- Optical Characterization Measurements of Commercial Polyvinyltoluene Scintillators for Accurate Geant4 Detector ModelingarXiv · 2026