Angle-dependent X-ray absorption correction in small- and wide-angle X-ray scattering: accounting for constrained scattering volume, beam profiles and capillary wall attenuation.

Jure Kovač, Andrej Jamnik, István Szilágyi, Matija Tomšič

Faculty of Chemistry and Chemical Technology University of Ljubljana Večna pot 113 Ljubljana1000 Slovenia. · MTA-SZTE Momentum Biocolloids Research Group, Department of Physical Chemistry and Materials Science, Interdisciplinary Research Center University of Szeged SzegedH-6720 Hungary.

Key points

From the paper's conclusion
  • The procedure extends path-length-based absorption correction to the Kratky-type geometry by averaging Beer-Lambert attenuation over the constrained illuminated scattering volume and by including experimentally measured primary beam profiles and capillary-wall absorption.

Abstract

A GPU-accelerated numerical method has been developed to calculate angle-dependent transmission factors A(2θ) for small- and wide-angle X-ray scattering (SWAXS) data collected with a Kratky-type camera using a line-collimated beam and a horizontal cylindrical capillary with a vertically oriented detector. The procedure extends path-length-based absorption correction to the Kratky-type geometry by averaging Beer-Lambert attenuation over the constrained illuminated scattering volume and by including experimentally measured primary beam profiles and capillary-wall absorption. Sensitivity analysis shows that the vertical illumination geometry dominates the calculated angle-dependent transmission factor, whereas the horizontal beam-width effect is negligible. Using measured beam profiles makes the correction instrument specific and avoids introducing an effective uniform beam height as an approximation. For strongly absorbing nona-fluoro-tert-butanol, the conventional scalar primary beam transmission correction fails after subtraction of empty-capillary scattering, while the numerical correction enables physically meaningful absolute-scale data reduction. Comparisons between independently calculated theoretical SWAXS intensities and the two reduced experimental datasets obtained sequentially using two different detectors in the same experiment on the same Kratky-type SWAXS camera provide a direct experimental cross check for the data-reduction workflow based on calculated transmission factors. They indicate that the observed limitation is associated with transmission measurement of a spectrally modified primary beam, rather than with the Kratky-type geometry itself. Importantly, the results show that the limitations of conventional scalar transmission corrections are not confined to extreme cases but may also affect routine water-based absolute-scale SWAXS calibration. A simple Beer-Lambert-type numerical scalar correction can still provide a useful qualitative approximation, whereas reliable quantitative analysis requires the full angle-dependent A(2θ) correction.

Read the paper