ABS and PETG as candidate materials for 3D-printed mammography phantoms: Physical and radiological characterization and assessment of TLD-indicated air kerma as a function of depth.
Crislane J Cesário, Greiciane J Cesário, Letícia L Campos Rodrigues, Linda V E Caldas, Maria da Penha Albuquerque Potiens
Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear (IPEN/CNEN), 05508-000, São Paulo, SP, Brazil. Electronic address: crislane.cesario@usp.br. · Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear (IPEN/CNEN), 05508-000, São Paulo, SP, Brazil; University of São Paulo (USP), Institute of Physics (IF), 05508-090, São Paulo, SP, Brazil.
Key points
From the paper's conclusion- Agreement between the physical and radiological properties and the depth measurements supports further investigation of these polymers in heterogeneous mammography phantom structures, although image-performance assessments are still required for validation.
Abstract
This study evaluated acrylonitrile butadiene styrene (ABS) and glycol-modified polyethylene terephthalate (PETG) printed by fused deposition modeling as candidate materials for mammography phantoms. ABS was printed at 90% infill and PETG at 100%. Characterization included determination of effective density, calculation of linear attenuation coefficients, and assessment of air kerma indicated by CaSO4:Dy dosimeters as a function of thickness and depth under the RQA-M2 and RQA-M4 radiation qualities. The dosimeter response over the range 0.2-10 mGy was fitted with a power function. The effective densities were 0.899 ± 0.002 g/cm3 for ABS and 1.169 ± 0.003 g/cm3 for PETG; at 28 keV, the respective linear attenuation coefficients were 0.203 ± 0.010 cm-1 and 0.308 ± 0.010 cm-1. The fit yielded a linearity index b = 1.07 ± 0.01, indicating a slightly supralinear thermoluminescent response, and the energy correction factor obtained for RQA-M4 was fc(E) = 0.931 ± 0.050. The surface-normalized profiles showed a progressive reduction in indicated air kerma, which was more pronounced for PETG and under RQA-M2. At 4 cm, the reduction ranged from 86.6% to 90.3% for PETG and from 78.3% to 82.3% for ABS under RQA-M2; under RQA-M4, it ranged from 77.0% to 81.4% and from 67.3% to 71.9%, respectively. Agreement between the physical and radiological properties and the depth measurements supports further investigation of these polymers in heterogeneous mammography phantom structures, although image-performance assessments are still required for validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Experimental determination of effective X-ray attenuation coefficients of 3D-printed materials under clinical mammography spectraFrontiers in Bioengineering and Biotechnology · 2026
- Experimental benchmarking of Geant4 electromagnetic physics lists for X-ray transmission and relative air-kerma profiles in 3D-Printed ABS and PETG mammography phantoms.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- X-Ray Attenuation Properties of Additive Manufacturing and 3D Printing Materials for Mimicking Tissues in Radiographic Phantoms Measured by CT from 70 to 140 kV: 2025 UpdateBiomimetics · 2026
- Investigation of the most suitable 3D printing method and tissue equivalent materials for mimicking glandular-adipose tissues for production of breast phantom.Physical and engineering sciences in medicine · 2026
- Characterization and 3D printing of a tissue equivalent anthropomorphic neck phantom for 2.5D dosimetric verification in radiotherapyJournal of Physics: Conference Series · 2026
- Development and validation of a 3D-printed dosimetry phantom for paediatric radiologyPhysical and Engineering Sciences in Medicine · 2026