Experimental determination of effective X-ray attenuation coefficients of 3D-printed materials under clinical mammography spectra
Adrián Belarra, Irene Hernández‐Girón, Julia Garayoa, Luis Carlos Martínez, Julio Valverde, María José Rot, Alejandro Ferrando, Antonio Fortes Martín,
Universidad Complutense de Madrid · University College Dublin
Key points
From the paper's conclusion- 3D-printed materials sharing the same polymer base can exhibit substantially different attenuation for DM/DBT spectra, resulting in different BR equivalences due to formulation variations (e.g., pigments and dopants). Therefore, phantom developers should not rely solely on base polymer designations and declared density when selecting materials for production of breast phantoms; instead, the attenuation of each specific formulation needs to be verified and periodically re-evaluated under the intended mammography conditions.
Abstract
Background 3D printing enables the fabrication of customized breast phantoms for image quality assessment in digital mammography (DM) and digital breast tomosynthesis (DBT). A major challenge is the absence of standardized, accessible methods to characterize the attenuation properties of 3D-printed materials under clinical DM/DBT spectra. Methods An experimental framework was implemented to determine the effective X-ray attenuation coefficient ( μ eff ) of six 3D-printed polymers (PLA, PET, resin, ABS, ABS+, HIPS) and reference breast tissue-equivalent materials (CIRS plates simulating different breast glandular/adipose ratios (BR) and PMMA) using two commercial DM/DBT systems, with and without anti-scatter grid. Step-wedges (0.5–5.5 cm) were imaged across multiple kVp and filter settings. The μ eff were obtained from measurements on images and fitted to an empirical model yielding μ 0 (attenuation at thickness tending to zero) and k (decay rate) to characterize beam hardening and scatter influences. 3D-reference material equivalences were evaluated based on μ eff and μ 0 . Results Beam hardening and scatter reduced μ eff with thickness, by 6%–14% with grid and 12%–28% without grid, with scatter contributing 47%–76% of the reduction in no-grid acquisitions. No significant differences were observed between the two mammography systems. Based on μ eff values, attenuation equivalences (within ±6%) were identified between 3D-printed and reference breast tissue-equivalent materials: PLA with BR 100/0; PET and resin with BR 70/30 and PMMA; ABS+ with BR 30/70 and BR 50/50. ABS and HIPS showed larger mismatches. The empirical model achieved excellent fits (R 2 > 0.99), with μ 0 values preserving attenuation ranking and enabling derivation of equivalent glandular proportions. Conclusion This framework demonstrates that routine clinical mammography systems can be used directly, without specialized instrumentation, to characterize 3D-printed materials as tissue surrogates. Several low-cost, widely available polymers were shown to reproduce breast tissue attenuation, supporting the local fabrication of anthropomorphic breast phantoms for realistic and clinically relevant image quality evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- ABS and PETG as candidate materials for 3D-printed mammography phantoms: Physical and radiological characterization and assessment of TLD-indicated air kerma as a function of depth.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Investigation of the most suitable 3D printing method and tissue equivalent materials for mimicking glandular-adipose tissues for production of breast phantom.Physical and engineering sciences in medicine · 2026
- X-Ray Attenuation Properties of Additive Manufacturing and 3D Printing Materials for Mimicking Tissues in Radiographic Phantoms Measured by CT from 70 to 140 kV: 2025 UpdateBiomimetics · 2026
- Experimental benchmarking of Geant4 electromagnetic physics lists for X-ray transmission and relative air-kerma profiles in 3D-Printed ABS and PETG mammography phantoms.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Verifying calculated dose to the breast in stereotactic radiotherapy: A practical phantom-based method.Journal of applied clinical medical physics · 2026
- Characterization and 3D printing of a tissue equivalent anthropomorphic neck phantom for 2.5D dosimetric verification in radiotherapyJournal of Physics: Conference Series · 2026