◆ Frontiers in Bioengineering and Biotechnology2026-02-05· Mammography

Experimental determination of effective X-ray attenuation coefficients of 3D-printed materials under clinical mammography spectra

Adrián Belarra, Irene Hernández‐Girón, Julia Garayoa, Luis Carlos Martínez, Julio Valverde, María José Rot, Alejandro Ferrando, Antonio Fortes Martín, Margarita Chevalier

一句话结论 · In one sentence

3D-printed materials sharing the same polymer base can exhibit substantially different attenuation for DM/DBT spectra, resulting in different BR equivalences due to formulation variations (e.g., pigments and dopants). Therefore, phantom developers should not rely solely on base polymer designations and declared density when selecting materials for production of breast phantoms; instead, the attenuation of each specific formulation needs to be verified and periodically re-evaluated under the intended mammography conditions.

原始摘要(英文原文)· Original abstract
Background 3D printing enables the fabrication of customized breast phantoms for image quality assessment in digital mammography (DM) and digital breast tomosynthesis (DBT). A major challenge is the absence of standardized, accessible methods to characterize the attenuation properties of 3D-printed materials under clinical DM/DBT spectra. Methods An experimental framework was implemented to determine the effective X-ray attenuation coefficient ( μ eff ) of six 3D-printed polymers (PLA, PET, resin, ABS, ABS+, HIPS) and reference breast tissue-equivalent materials (CIRS plates simulating different breast glandular/adipose ratios (BR) and PMMA) using two commercial DM/DBT systems, with and without anti-scatter grid. Step-wedges (0.5–5.5 cm) were imaged across multiple kVp and filter settings. The μ eff were obtained from measurements on images and fitted to an empirical model yielding μ 0 (attenuation at thickness tending to zero) and k (decay rate) to characterize beam hardening and scatter influences. 3D-reference material equivalences were evaluated based on μ eff and μ 0 . Results Beam hardening and scatter reduced μ eff with thickness, by 6%–14% with grid and 12%–28% without grid, with scatter contributing 47%–76% of the reduction in no-grid acquisitions. No significant differences were observed between the two mammography systems. Based on μ eff values, attenuation equivalences (within ±6%) were identified between 3D-printed and reference breast tissue-equivalent materials: PLA with BR 100/0; PET and resin with BR 70/30 and PMMA; ABS+ with BR 30/70 and BR 50/50. ABS and HIPS showed larger mismatches. The empirical model achieved excellent fits (R 2 > 0.99), with μ 0 values preserving attenuation ranking and enabling derivation of equivalent glandular proportions. Conclusion This framework demonstrates that routine clinical mammography systems can be used directly, without specialized instrumentation, to characterize 3D-printed materials as tissue surrogates. Several low-cost, widely available polymers were shown to reproduce breast tissue attenuation, supporting the local fabrication of anthropomorphic breast phantoms for realistic and clinically relevant image quality evaluation.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Experimental determination of effective X-ray attenuation coefficients of 3D-printed materials under clinical mammography spectra — 科研速览 Science Skim