Analysis of an irradiated uranium sample for source attribution without chemical separation using microplasma ionization and ultrahigh resolution mass spectrometry
Joseph V. Goodwin, Titeana L. M. Cook, Suraj Shrestha, Craig M. Marianno, Benjamin T. Manard, C. M. Folden, R. Kenneth Marcus
Clemson University · Texas A&M University
Abstract
Abstract The use of element isotope ratios has great potential in not only determining the reactor type used to produce plutonium (Pu) but also in determining the burnup and the time since irradiation. While a powerful nuclear forensic technique, determining element isotope ratios is complicated by severe isobaric interferences when performed on typical inductively coupled plasma mass spectrometers. Such analyses require extensive chemical separations prior to analysis to alleviate the inter-elemental isobars. Ultrahigh mass resolution spectrometry provides a potential alternative, greatly reducing the complexity of sample preparation and turnaround times for these critical measurements. To demonstrate the power of the approach, a sample of irradiated, depleted uranium was analyzed with the liquid sampling—atmospheric pressure glow discharge ion source coupled to an Orbitrap mass spectrometer. The Orbitrap is augmented with an external data acquisition system, Spectroswiss’s FTMS-Booster X2T, allowing collection of extended ion transients, providing higher mass resolution. In using this approach, the 150 Sm/ 149 Sm and 152 Sm/ 149 Sm isotope ratios were found to be within 20% of predicted values without any chemical separations and without mass bias corrections. In addition, the 240 Pu/ 239 Pu isotope ratio was determined, free from the 238 UH + interferences common to the ICP-MS platforms, while at the same time allowing for the determination of U isotopic signatures. While these demonstrative results are from a single sample, the advantages of the microplasma/ultrahigh mass resolution approach to intra-element isotope ratio determinations are clear.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance evaluation of MAGIX-UPu software for determination of uranium and plutonium isotopic composition with medium resolution gamma-ray detectors.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Calibration of the NDUSRM series and further measurements of 236U/238U isotopic ratios in the Ewing collection.Journal of environmental radioactivity · 2026
- Single-Particle Isotopic Analysis Using the Liquid Sampling-Atmospheric Pressure Glow Discharge Microplasma Coupled to an Orbitrap Mass SpectrometerAnalytical Chemistry · 2026
- Validation of LG-SIMS analysis of plutonium-bearing particles using Pu-impregnated silica particles and its application to the Daigo Fukuryu-Maru sample.Analytical methods : advancing methods and applications · 2026
- Multi-Collector Extreme Ultraviolet Laser Mass Spectrometry (MC-EUV-MS): Using High-Energy Photoionization for High-Spatial-Resolution Isotope Ratio Mapping of Actinide Materials.Analytical chemistry · 2026
- Capsule-Type Hydrothermal Extraction-Coupled NanoESI-MS/MS for Analysis of Uranium Isotope in Sub-milligram Lunar Regolith Samples.Analytical chemistry · 2026