Multi-Collector Extreme Ultraviolet Laser Mass Spectrometry (MC-EUV-MS): Using High-Energy Photoionization for High-Spatial-Resolution Isotope Ratio Mapping of Actinide Materials.
Lydia A Rush, Matthew A Coble, Carmen S Menoni, Andrew M Duffin
Pacific Northwest National Laboratory , Richland, Washington99354, United States. · Department of Electrical and Computer Engineering, Colorado State University, Fort Collins, Colorado80523, United States.
Key points
From the paper's conclusion- We demonstrate the high spatial resolution of the MC-EUV-MS by mapping uranium oxide microparticles using 500 nm ablation spots.
Abstract
Isotope maps are typically collected with a multi-collector mass spectrometer (MC-MS) coupled to a high-spatial-resolution ionization source, yet few ionization sources compatible with MC-MS systems can operate robustly at the submicron scale. To meet this need, we couple an extreme ultraviolet (EUV) laser photon-based ionization source to a modified Thermo Scientific NeptunePlus MC-MS by removing and replacing the inductively coupled plasma (ICP) with the EUV laser system, herein called the MC-EUV-MS. The 46.9 nm EUV laser light is focused to sub-micrometer spots for sample ablation while simultaneously realizing high-energy photoionization using 26.4 eV photons. Thus, the EUV probe combines ablation and ionization in a single pulse. We characterize the newly built MC-EUV-MS by analyzing different actinide-bearing materials. We achieve a sample utilization efficiency (SUE) of 0.006% for 238U using the NIST 610 and 612 glasses at a low +2 kV extraction voltage for accelerating ions off the sample. Higher extraction voltages up to +5 kV point to increased ion yields. We measure 235U/238U isotope ratios from three uranium samples ranging from natural to low enriched. We demonstrate the high spatial resolution of the MC-EUV-MS by mapping uranium oxide microparticles using 500 nm ablation spots. Compared to isotope maps collected over the same sample area using large geometry secondary ion mass spectrometry, the MC-EUV-MS shows significantly higher spatial resolution within the scope of this study.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Capsule-Type Hydrothermal Extraction-Coupled NanoESI-MS/MS for Analysis of Uranium Isotope in Sub-milligram Lunar Regolith Samples.Analytical chemistry · 2026
- Analysis of an irradiated uranium sample for source attribution without chemical separation using microplasma ionization and ultrahigh resolution mass spectrometryJournal of Radioanalytical and Nuclear Chemistry · 2026
- High conversion efficiency of extreme ultraviolet light generation at 11 nm from a laser-produced Xe plasmaJournal of Physics B: Atomic, Molecular and Optical Physics · 2026
- Initial Performance Evaluation of an Ultraviolet 157 nm Laser Ablation System for ICP-MS Analysis.Analytical chemistry · 2026
- Data‐driven elemental descriptors for rational design of high‐sensitivity extreme ultraviolet photoresistsSmart Molecules · 2026
- Calibration of the NDUSRM series and further measurements of 236U/238U isotopic ratios in the Ewing collection.Journal of environmental radioactivity · 2026