Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine

Performance evaluation of MAGIX-UPu software for determination of uranium and plutonium isotopic composition with medium resolution gamma-ray detectors.

I Espagnon, A-C Simon

Université Paris-Saclay, CEA, List, Palaiseau, F-91120, France. Electronic address: isabelle.espagnon@cea.fr. · Université Paris-Saclay, CEA, List, Palaiseau, F-91120, France.

Key points

From the paper's conclusion
  • The results obtained for uranium spectra are also satisfying, with an average total uncertainty of 8.9% for 235U.

Abstract

This study evaluates the performance of the MAGIX-UPu software for determining the isotopic composition of uranium (U) and plutonium (Pu) using medium-resolution gamma-ray detectors in nuclear industry applications. The analysis involves two Lanthanum Bromide (LaBr3) detectors of different sizes and an H3D gamma camera (Cadmium Zinc Telluride, CZT). Developed by CEA List, this code performs a comprehensive and automatic spectrum analysis, including calibration, peak deconvolution, and simultaneous determination of relative efficiency and isotopic composition. The analysis revealed that the H3D gamma camera provided significantly better results due to its superior energy resolution. The mass fractions of 239Pu can notably be obtained with an average total uncertainty of 1.3%. The results obtained for uranium spectra are also satisfying, with an average total uncertainty of 8.9% for 235U. With the LaBr3 detectors, MAGIX-UPu can determine the isotopic composition of U and Pu, but the poor resolution complicates the analysis and makes consistency checks difficult. Nevertheless, the results provided by MAGIX-UPu for plutonium are good compared to other software.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers