Single-Particle Isotopic Analysis Using the Liquid Sampling-Atmospheric Pressure Glow Discharge Microplasma Coupled to an Orbitrap Mass Spectrometer
Suraj Shrestha, Sarah E. Szakas, Hunter B. Andrews, Benjamin T. Manard, R. Kenneth Marcus
Clemson University · Oak Ridge National Laboratory
Key points
From the paper's conclusion- Presented here is the first demonstration of an ultrahigh resolution method for single particle (SP) isotopic analysis using the liquid sampling-atmospheric pressure glow discharge (LS-APGD) microplasma ionization source coupled to an Orbitrap mass spectrometer and FTMS Booster X2T acquisition/processing system.
Abstract
Isotope ratio (IR) determinations on individual particles can provide valuable information relative to the sample, including processing and dating, which could be valuable to geological and nuclear material analysis communities. In comparison to “bulk” measurements, in which all particle information is lost and homogenized within a sample, “particle” measurements allow for fingerprinting and can provide unique insights related to the sample’s nano- and micro- compositions. Furthermore, the complex nature of particles, with potential isobaric interferences from either the matrix or other elements within the same particle, poses a significant analytical challenge for conventional mass spectrometry platforms employed for elemental analysis due to their limited mass resolution. Presented here is the first demonstration of an ultrahigh resolution method for single particle (SP) isotopic analysis using the liquid sampling-atmospheric pressure glow discharge (LS-APGD) microplasma ionization source coupled to an Orbitrap mass spectrometer and FTMS Booster X2T acquisition/processing system. For proof of concept, well-characterized CeO 2 microparticles with a nominal diameter of 1.0 μm were analyzed at a mass resolution of ∼330,000. Important instrumental parameters were optimized to enable the detection of single particles. The 142 Ce/ 140 Ce ratio of individual particles and the population average were determined and compared to an SP inductively coupled plasma time-of-flight mass spectrometry (ICP-TOF-MS) analysis. The isotopic accuracy and precision were determined by two methods and found to be in good agreement with the SP-ICP-TOF-MS method, with the linear regression slope method providing a more accurate ratio, showing a ∼1.4% relative difference from the ratio determined from a particle digest. The encouraging performance of the method was further supported by a determined detection limit of 2.9 fg ( 142 Ce). The developed method is anticipated to overcome many of the challenges posed by isobaric interferences encountered in conventional mass spectrometry techniques when analyzing real-world samples for particle isotopic composition.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- IsotopeTrack: a fast and flexible application for the analysis of SP-ICP-TOF-MS datasetsEnvironmental Chemistry · 2026
- All-Laser Suite Analytical Platform (Microsampling, Catapulting, Trapping, and LIBS) for the Characterization of Discrete Domains in Samples of Astrochemical Interest.Analytical chemistry · 2026
- Analysis of an irradiated uranium sample for source attribution without chemical separation using microplasma ionization and ultrahigh resolution mass spectrometryJournal of Radioanalytical and Nuclear Chemistry · 2026
- Single particle - MC-ICP-MS for isotopic analysis of uranium particles.Analytical methods : advancing methods and applications · 2026
- Stable Carbon Isotope Analysis of PFOA and PFOS Using Orbitrap Mass SpectrometryEnvironmental Science & Technology Letters · 2026
- δ13C Determination of Benzoic Acid by High-Resolution Orbitrap MS: How Sample Introduction Strategy Governs Accuracy and Precision.Journal of the American Society for Mass Spectrometry · 2026