Capsule-Type Hydrothermal Extraction-Coupled NanoESI-MS/MS for Analysis of Uranium Isotope in Sub-milligram Lunar Regolith Samples.
Lili Song, Yongpeng Liu, Chun Wan, Xinglei Zhang, Zhaobin Yan, Guoqi Liu, Qingfei Wang, Fengjun Nie,
Jiangxi Key Laboratory for Mass Spectrometry and Instrumentation, East China University of Technology, Nanchang330013, P. R. China. · State Key Laboratory of Nuclear Resources and Environment, East China University of Technology, Nanchang330013, P. R. China.
Key points
From the paper's conclusion- Application of the CTHE-nanoESI-MS/MS method to Chang'e-5 lunar regolith afforded a uranium concentration of 1.44 ± 0.06 μg/g and a 235U/238U ratio of 0.007198 ± 0.000102 (RSD = 1.42%, δ = -8.13‰ relative to the certified natural uranium value).
Abstract
Uranium concentration and isotopic data from returned extraterrestrial materials provide constraints on planetary differentiation, thermal evolution, and chronological frameworks. However, the extreme paucity of such specimens demands analytical methodologies capable of analysis at ultratrace sample amounts. Herein, we report an integrated workflow coupling capsule-type hydrothermal extraction (CTHE) with nanoelectrospray ionization tandem mass spectrometry (nanoESI-MS/MS) for the determination of uranium isotope in sub-milligram lunar regolith. Leveraging the hydrostatic pressure balancing strategy, sub-milligram samples were hydrothermally digested within a polytetrafluoroethylene micro-capsule using HF/HNO3 at high-temperature and high-pressure conditions. Following controlled elimination of residual acids, uranium was converted to the [UO2(NO3)3]- complex and analyzed by nanoESI-MS/MS in the negative ion mode via selective reaction monitoring of the characteristic transitions m/z 456.0 → 410.0 (238U) and m/z 453.0 → 407.0 (235U). The optimized method affords a limit of detection of 0.63 μg/L, a linear dynamic range spanning 0.01-2.00 mg/L (R2 > 0.99), a sample consumption as low as 0.1 mg, and spike recoveries of 92.4-97.4% with relative standard deviations (RSDs) < 4% (n = 3). Application of the CTHE-nanoESI-MS/MS method to Chang'e-5 lunar regolith afforded a uranium concentration of 1.44 ± 0.06 μg/g and a 235U/238U ratio of 0.007198 ± 0.000102 (RSD = 1.42%, δ = -8.13‰ relative to the certified natural uranium value). Notably, this approach reduces sample consumption by two orders of magnitude relative to conventional hydrothermal digestion combined ICP-MS detection methods. This methodology establishes an analytical paradigm for uranium isotope geochemistry in sample-limited materials.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of a Flow-through Cell for Ultrasonic Extraction (UE)─Single Particle (SP)─ICP-MS─an Approach for Nano/Microparticle Elemental and Isotopic Analysis.Analytical chemistry · 2026
- Photothermal activation by hollow multishelled structure for efficient uranium extractionNational Science Open · 2026
- Measurement of the Isotopic Composition of Uranium in Geological Samples by MC-ICP-MS Using a TOPO-Based Extraction Chromatographic ProtocolAnalytical Chemistry · 2026
- Multi-Collector Extreme Ultraviolet Laser Mass Spectrometry (MC-EUV-MS): Using High-Energy Photoionization for High-Spatial-Resolution Isotope Ratio Mapping of Actinide Materials.Analytical chemistry · 2026
- Light-Enhanced Uranium Capture by TiO2–Incorporated Polyamidoxime Nanofibers through Adsorption and PhotoreductionACS Applied Polymer Materials · 2026
- Hydrogen‐Bond‐Assisted Secondary Amine‐Phosphonate Tandem Sites on Mesoporous Nano‐Silica for Uranyl Carbonate Capture From Alkaline WastewaterSmall · 2026