Monte Carlo-based secondary dose check platform for the world's first ultra-compact upright proton therapy system
Toghrul Alakbarzade, Emir Musa Olgun, Yuan James Rao, Billy W Loo Jr, Adam Johansson, Yong Yang, Murat Surucu, Mehmet Burcin Unlu,
Key points
From the paper's conclusion- This work aimed to develop and validate an independent Monte Carlo (MC) secondary dose check of the RayStation treatment planning system (TPS) for the Mevion S250-FIT, the world's first ultra-compact upright proton therapy system.
Abstract
Objective. This work aimed to develop and validate an independent Monte Carlo (MC) secondary dose check of the RayStation treatment planning system (TPS) for the Mevion S250-FIT, the world's first ultra-compact upright proton therapy system. Approach. The treatment nozzle, including the energy modulation system of boron carbide and polycarbonate plates and the adaptive aperture, was modeled in the TOPAS (TOol for PArticle Simulation) MC code, and the beam model was commissioned against integrated depth dose curves, in-air spot profiles, and dose per monitor unit measured at 12 energy settings. Based on this model, an in-house converter, RayStation2TOPAS, recalculates the TPS plans by reconstructing each pencil beam spot, the aperture for each energy layer, and the seated patient's rotation. In total, 15 water phantom plans and 25 clinical plans were recalculated and compared with the TPS by the point dose, with a tolerance of 5%, and by three-dimensional gamma analysis at 3%/2mm, with a pass rate of at least 90%. Main results. The simulated proton range agreed with the measurement within 0.25mm, the spot full width at half maximum within 6.8%, and the dose per monitor unit within 1.3%. The point dose differed from the TPS by -0.76% to +1.98% in the water phantom plans and by -3.03% to +3.99% in the clinical plans, with gamma pass rates of 96.39% to 99.93% and 90.08% to 99.98%, so every plan passed both tests. Significance. To our knowledge, this is the first independent MC secondary dose check for the S250-FIT, and it complements measurement-based patient-specific quality assurance of its plans.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Proton treatment-plan dose verification with a DPM-based Monte Carlo engine.Frontiers in oncology · 2026
- Design, implementation, and assessment of an independent second-check system for proton and carbon-ion pencil-beam scanning: delivery-aware Monte Carlo and microdosimetric RBE-weighted dose.Frontiers in oncology · 2026
- Monte Carlo simulations as the cornerstone of next-generation patient-specific quality assurance in particle therapyFrontiers in Oncology · 2026
- Experimental verification of clinical dose and LET maps for LET-informed proton therapy.Physics in medicine and biology · 2026
- Configuration of a Monte Carlo-Based Secondary Plan Verification Tool: An Analytical Approach to User-Specific Parameter Tuning.Biomedical physics & engineering express · 2026
- RadCalc 3D Monte-Carlo: tuning and independent dosimetric validation of radiotherapy treatment plans.Biomedical physics & engineering express · 2026