Biomedical physics & engineering express

RadCalc 3D Monte-Carlo: tuning and independent dosimetric validation of radiotherapy treatment plans.

Francesco Caridi, Giuseppe Paladini, Jacopo Piccione, Domenico Majolino, Bruno M Itri, Salvatore Pittera, Giuseppe Sceni, Valentina Venuti

Dipartimento di Scienze Matematiche e Informatiche, Scienze Fisiche e Scienze della Terra, Università degli Studi di Messina, V.le F. Stagno D'Alcontres 31, 98166 Messina, Italy. · Dipartimento di Scienze Biomediche, Odontoiatriche e delle Immagini Morfologiche e Funzionali, Università degli Studi di Messina, c/o A.O.U. Policlinico 'G. Martino'Via Consolare Valeria 1, 98125 Messina, Italy.

Observational

Key points

From the paper's conclusion
  • Objective.This work aimed at implementing a new recalculation software, 'RadCalc' (LAP GmbH Laser Applikationen), independent of the monitor units (MUs) and dosimetric verification of the radiotherapy treatment plans specific for each patient, using the Monte-Carlo (MC) algorithm.Approach.The experimental procedure for the simulation of the treatment plans by the software was started by measuring

Abstract

Objective.This work aimed at implementing a new recalculation software, 'RadCalc' (LAP GmbH Laser Applikationen), independent of the monitor units (MUs) and dosimetric verification of the radiotherapy treatment plans specific for each patient, using the Monte-Carlo (MC) algorithm.Approach.The experimental procedure for the simulation of the treatment plans by the software was started by measuring dosimetric parameters of the operating machine (Elekta Versa HD linear accelerator), subsequently imported into the software. After that, the software tuning on the treatment planning system Monaco v. 5.53 (Elekta) was carried out, which allowed to extract the most suitableAdditional Radiation to Light Field Offsetparameter to obtain the best performing MC algorithm model for the simulation of the treatment plans. Finally, the above-mentioned model was validated through a statistical analysis comparing the plans recalculated by RadCalc and those experimentally verified on the OCTAVIUS 4D phantom (PTW) and Octavius 1500 (PTW) Detector.Main results.The obtained results showed a good statistical agreement both in the tuning and in the validation part of the MC algorithm model used. The effectiveness of using the RadCalc software was further demonstrated by the high accuracy of the recalculation of the MUs.Significance.It should be noted that the secondary recalculation of the dose, in accordance with Italian legislation on patient radiation protection is one of the actions that the medical physics specialist must'prepare for the evaluation and correct verification of the doses administered to patients'. In this context, it has to be emphasized that a multicenter study group is being set up, within the Italian Association of Medical and Health Physics (AIFM), which brings together the users of RadCalc, aimed at exchanging experiences and improving the performance of the aforementioned software.

Read the paper