Frontiers in oncology

Design, implementation, and assessment of an independent second-check system for proton and carbon-ion pencil-beam scanning: delivery-aware Monte Carlo and microdosimetric RBE-weighted dose.

Jun Tan, Keith Furutani, Alessio Parisi, Shannon Hartzell, Chris Beltran

Department of Radiation Oncology, Mayo Clinic Florida, Jacksonville, FL, United States.

Computational

Key points

From the paper's conclusion
  • Conclusion A common delivery-aware platform was implemented to support independent physical dose and microdosimetric RBE-weighted dose assessment for proton and carbon-ion treatments while retaining beam-specific dose conventions and biological model information.

Abstract

BACKGROUND: Independent second checks for proton and carbon-ion pencil-beam scanning should assess the planned physical dose, the expected delivery sequence, and the model-dependent relative biological effectiveness (RBE)-weighted dose. OBJECTIVE: The aim of this study was to design, implement, and assess a traceable second-check system that combines delivery-aware Monte Carlo transport with microdosimetric RBE calculation for both particle species. METHODS: The system receives Digital Imaging and Communications in Medicine (DICOM) planning data, converts the discrete spot sequence into a predicted delivery sequence for dose-driven continuous scanning (DDCS), generates weighted particle histories, and performs independent TOPAS/Geant4 transport. It scores physical dose, dose-averaged linear energy transfer (LETd), and voxel-level microdosimetric quantities. For both beam types, the scored lineal-energy distribution is represented using abridged microdosimetry distribution methodology (AMDM) and processed with the Mayo Clinic Florida microdosimetric kinetic model (MCF MKM). The proton branch additionally produces a fixed-RBE-weighted dose using RBE = 1.1. No fixed-RBE model is used for carbon ions. Proton assessment included component tests, single-energy source, transport, and scoring checks at 70.0, 170.2, and 229.1 MeV, patient-level evaluations using de-identified plans, and operational timing. Carbon-ion assessment used a pre-existing Monte Carlo dataset to evaluate component-level handling and visualization of AMDM descriptors, the MCF MKM RBE factor, and RBE-weighted dose. RESULTS: The integrated workflow completed plan receiving, delivery-sequence generation, particle transport, physical and biological dose calculation, comparison, and report generation. The proton case for which detailed quantitative end-to-end results were reported was classified as REVIEW REQUIRED because not all dosimetric comparison criteria were met. A pre-existing carbon-ion dataset was used to verify the component-level handling and separate presentation of physical dose, AMDM descriptors, RBE factor, and RBE-weighted dose. Predicted DDCS timing and the stationary-versus-moving irradiation components were not benchmarked against machine logs or measurements. CONCLUSION: A common delivery-aware platform was implemented to support independent physical dose and microdosimetric RBE-weighted dose assessment for proton and carbon-ion treatments while retaining beam-specific dose conventions and biological model information. Further local commissioning and multicase validation are required before clinical use.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers