Experimental verification of clinical dose and LET maps for LET-informed proton therapy.
Mads Lykke Jensen, Michelle Van Heerden, Liliana Stolarczyk, Michele Togno, Francesca Albertini, Niels Bassler, Anne Vestergaard, Arturs Meijers,
Department of Radiation Safety and Security, Paul Scherrer Institut PSI, Forschungsstrasse 111, OFLD/005, Villigen, Villigen, 5232, Switzerland. · Center for Proton Therapy, Paul Scherrer Institute PSI, Forschungsstrasse 111, Villigen, Aargau, 5232, Switzerland.
Key points
From the paper's conclusion- Given this and the system’s ability to simultaneously measure 2D dose and LET D maps with high spatial resolution and minimal field perturbation, the system shows promise for future commissioning workflows for LET-informed proton therapy.
Abstract
Objective:Linear energy transfer (LET)-informed proton therapy has the potential to improve biological treatment planning beyond the conventional, constant-RBE assumption. Its clinical translation, however, requires experimental tools capable of verifying spatially resolved dose and dose-averaged LET (LETD) distributions under clinically relevant conditions. We recently introduced an optically stimulated luminescence (OSL) film-based detector system that simultaneously measures two-dimensional (2D) absorbed dose and LETD. This work provides a first assessment of the system for verifying dose and LETDdistributions from clinical proton therapy plans. Approach:The system was tested at two proton therapy centers, ranging from simple uniform fields to more complex geometries, including Bragg curve measurements and a clinically realistic anthropomorphic head phantom treatment plan. Measured dose and LETDdistributions were compared against commercial treatment planning system (TPS) calculations and, for selected fields, to external Monte Carlo particle transport simulations due to the absence of traceability of LETD. For non-uniform irradiations, agreement was assessed using gamma analysis with 3%/2 mm and 5%/2 mm criteria for dose and LETD, respectively. Main results:For uniform fields, measured dose and LETDdistributions generally agreed with TPS and independent Monte Carlo simulations across a range of clinically relevant irradiation conditions. In the Bragg curve measurement, gamma pass rates were 96.6% for dose and 99.7% for LETD, with the largest dose discrepancy observed near the Bragg peak, consistent with a current over-correction of detector quenching in high-LET regions to be improved in future work. In the anthropomorphic phantom, gamma pass rates ranged from 75% to 100% for dose and from 45% to 95% for LETD, with a single layer showing a sub-50% pass rate for LETD. Failing regions were confined to high-dose superior layers for dose and distal-edge, high-LET regions for LETD. Significance:The presented results support the detector system's use as an experimental benchmarking tool for TPS-calculated dose and LETDmaps. Given this and the system's ability to simultaneously measure 2D dose and LETDmaps with high spatial resolution and minimal field perturbation, the system shows promise for future commissioning workflows for LET-informed proton therapy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Interpreting Dose×LET volume histograms using difference mapping in Spot Delete proton planningJournal of Applied Clinical Medical Physics · 2026
- Estimation of dose and linear energy transfer for multi-spot single-energy proton probing beams using in-beam PET and neural network.Physics in medicine and biology · 2026
- Monte Carlo-based secondary dose check platform for the world's first ultra-compact upright proton therapy systemarXiv · 2026
- Estimation of LET distributions ofα-particles from measured scintillator-based trajectory images.Physics in medicine and biology · 2026
- Quality assurance for deep learning-based proton therapy planning in routine clinical use for oropharyngeal cancer.Physics and imaging in radiation oncology · 2026
- A cascade deep learning network for simultaneous prediction of dose and linear energy transfer in proton therapy of prostate cancer.Physics in medicine and biology · 2026