In vivo Dose Measurement in Cervical Cancer Brachytherapy Using Dysprosium-doped Calcium Sulfate Thermoluminescent Dosimeters: A Phantom Study.
Prabhat Krishna Sharma, Saral Kumar Gupta, Devesh Gupta
Department of Physical Sciences, Banasthali Vidyapith, Tonk, Rajasthan, India. · Department of Radiological Physics, Dr. S. N. Medical College, Regional Cancer Institute, Jodhpur, Rajasthan, India.
Key points
From the paper's conclusion- The results confirm that with proper calibration, CaSO4:Dy Teflon TLD discs could also be a useful and trustworthy tool for independent dose verification in HDR BT. This approach can enhance routine quality assurance practices and improve confidence in dose delivery accuracy.
Abstract
AIMS: High-dose-rate (HDR) brachytherapy (BT) is an important part of the treatment of cervical cancer. Although there have been advances in three-dimensional image-guided treatment planning systems (TPSs), uncertainties in applicator positioning and anatomical variations can produce discrepancies in planned and delivered doses. This study aimed to evaluate the clinical feasibility and reliability of dysprosium-doped calcium sulfate (CaSO4:Dy) Teflon thermoluminescent dosimeters (TLDs) for independent dose verification in Co-60 HDR cervical BT. MATERIALS AND METHODS: For calibration, the CaSO4:Dy Teflon TLD discs were held at a 5-cm depth in a phantom to closely mimic the conditions of dose measurements in a phantom. We specially constructed a phantom compatible with vaginal cylinders and Fletcher-Suit applicators to place TLDs at clinically significant sites. The measured doses were compared with TPS-calculated values by uncertainty analysis and statistical evaluation. RESULTS: In-depth phantom measurements demonstrated that, for the same delivered dose, the response of CaSO4:Dy TLDs at 1-cm and 5-cm depth in the phantom differed by only about 4%. The TLD system demonstrated a combined uncertainty of 11.80% (c = 2), with absolute mean deviation ranging from 6% to 9% across all measurement points. Bland-Altman analysis demonstrates that more than 95% of the paired observations were distributed within the limit. CONCLUSION: The results confirm that with proper calibration, CaSO4:Dy Teflon TLD discs could also be a useful and trustworthy tool for independent dose verification in HDR BT. This approach can enhance routine quality assurance practices and improve confidence in dose delivery accuracy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR BrachytherapyarXiv · 2026
- Preliminary feasibility results for implementing a phantom model for external auditing of HDR brachytherapy with an Ir-192 HDR source: an IAEA collaborative project.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Validation of a detector-response correction model for in vivo dosimetry in high-dose-rate prostate brachytherapy.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Dosimetric analysis along core-needle biopsies in prostate HDR brachytherapy.Physics in medicine and biology · 2026
- Investigation of TG-43 dose calculation algorithm for volumetric HDR Brachytherapy using Monte Carlo simulation in TOPAS.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Development of a low-cost 3D-printed phantom for end-to-end testing of model-based dose calculation algorithms (MBDCAs) in high-dose-rate (HDR) brachytherapy.Journal of applied clinical medical physics · 2026