Analytical Agreement and Diagnostic Performance of a Commercial Point-of-Care Lipid Testing System for Lipid Profile and Atherogenic Ratios.
Susana Guadalupe Guzmán-Aquino, Roberto Rivera-Sánchez, Karla Aidee Aguayo-Cerón, Manuel Solis-Navarro, Claudia Camelia Calzada-Mendoza, Rodrigo Romero-Nava, Rodolfo Pinto-Almazán, Adriana Flores-Miranda,
Laboratorio de Modelación Bioestadística, Sección de Estudios de Posgrado e Investigación, Escuela Superior de Medicina, Instituto Politécnico Nacional, Plan de San Luis y Díaz Mirón s/n, Mexico City 11340, Mexico. · División de Investigación, Hospital Juárez de México, Av. Instituto Politécnico Nacional, Mexico City 07760, Mexico.
Key points
From the paper's conclusion- Objective: This study aimed to evaluate the agreement, systematic bias, and diagnostic performance of the Mission® Cholesterol 3-in-1 system for lipid profile measurements and atherogenic ratios.
Abstract
Background: Point-of-care testing (POCT) devices offer a rapid alternative for lipid assessment, but comprehensive analytical validation against reference laboratory methods remains limited. Objective: This study aimed to evaluate the agreement, systematic bias, and diagnostic performance of the Mission® Cholesterol 3-in-1 system for lipid profile measurements and atherogenic ratios. Methods: A prospective cross-sectional agreement study was conducted among 120 participants. Lipid profiles were measured using eleven Mission® Cholesterol 3-in-1 analyzers and compared with those measured using the reference laboratory method. Agreement was assessed using the intraclass correlation coefficient (ICC), concordance correlation coefficient (CCC), coefficient of determination (R2), root mean square error (RMSE), mean absolute error (MAE), percentage bias, Bland-Altman plots, Passing-Bablok regression, and diagnostic accuracy. Results: Triglycerides demonstrated excellent agreement (ICC = CCC = 0.972; bias = 5.19%), whereas total cholesterol, HDL-C, non-HDL-C, and LDL-C showed good agreement (ICC range: 0.817-0.854). Among the atherogenic ratios, TG/HDL-C (ICC = 0.932) and TG/LDL-C (ICC = 0.931) exhibited the highest concordance. Passing-Bablok regression revealed constant bias for all analytes, except for LDL-C estimated using the Sampson equation and the TG/LDL-C ratio, while proportional bias was observed for all analytes, except for triglycerides and the TG/LDL-C ratio. Despite these systematic differences, diagnostic performance remained excellent, with sensitivities of 94.29-98.33% and specificities of 68.33-91.80%. Conclusion: The Mission® Cholesterol 3-in-1 system shows good analytical agreement with laboratory testing, particularly for triglycerides and the HDL-C, TG/HDL-C, and TG/LDL-C ratios. Although systematic bias limits complete interchangeability with reference methods, its excellent diagnostic performance supports its use as a rapid point-of-care screening tool, especially in resource-limited settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative Performance Analysis of Three Point-of-Care Glucose Testing Devices against a Central Laboratory Reference Method.EJIFCC · 2026
- Assessing the reliability and validity for metabolic syndrome diagnosis using point-of-care analyzers in a mobile setup - a pilot study.BMC primary care · 2026
- A-038 Validation of a Specialized Lipids Panel LDT for Serum Analysis Using the Roche Cobas 8000 c702 PlatformClinical Chemistry · 2026
- Portable Home‐Use Triglyceride Meter Demonstrates Good Agreement With Plasma Laboratory Measures During Pregnancies Complicated by Metabolic DiseaseEndocrinology Diabetes & Metabolism · 2026
- B-015 Precision Compatibility - Clinical Comparative Study of Lipid Panel Reagents on Six Mainstream InstrumentsClinical Chemistry · 2026
- Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation.Indian journal of community medicine : official publication of Indian Association of Preventive & Social Medicine · 2026