Functionality status and drivers of non-functionality of the standardised laboratory analysers in Tanzania mainland: Findings from a National Equipment Harmonisation and Standardisation Mapping Exercise
P. Torokaa, R. Makala, J. D. Kitukulu, E. M. Shija, M. E. Kelly, G. E. Kwavava, R. S. Mkala, I. G. Paul,
Key points
From the paper's conclusion- This study determined the functionality status of standardised laboratory analysers in Tanzania mainland and identified the drivers of non-functionality.
Abstract
Introduction Reliable, well-maintained laboratory equipment is essential for timely, accurate and reliable laboratory results for patient care, surveillance, and health system resilience. Tanzania implemented the Standard Medical Laboratory Equipment Guideline (SMLEG-2018) to harmonise procurement and improve efficiency across a tiered laboratory system and network. However, equipment functionality gaps remain a critical challenge for healthcare delivery. This study determined the functionality status of standardised laboratory analysers in Tanzania mainland and identified the drivers of non-functionality. Methods We conducted a facility-based cross-sectional study to identify the causes of non-functionality of standardised chemistry, haematology, and urine analysers across mainland Tanzania. We assessed 3,044 analysers and used Pareto analysis to identify the primary drivers of non-functionality. Results Of the 3,044 standardised analysers assessed, 987 (32.4%) chemistry 1,308 (43.0%) haematology, and 749 (24.6%) urine) assessed; 397 (13.0%) were non-functional. Proportionally, urine analysers were the most non-functional (n=176/749, 23.5%), followed by chemistry (n=135/987, 13.7%), and haematology (n=86/1308, 6.6%) analysers, p<0.001. Urine analysers had significantly higher non-functional equipment (23.5%), followed by Chemistry analysers (13.7%), than Haematology analysers (6.6%), p<0.001. Equipment breakdown was the leading cause of non-functionality, accounting for 70.9% (61/86), 70.5% (124/176) and 60.0% (81/135) of the non-functional haematology, urine and chemistry analysers, respectively, among all causes identified for each analyser type. Conclusion Although the SMLEG 2018 implementation identified broad coverage of functional standardised laboratory equipment, persistent equipment breakdowns of more than 10% highlight a systemic, concerning gap that requires immediate corrective action to sustain high performance. Strengthening preventive maintenance, timely repair services and sustainable equipment lifecycle planning is critical to sustaining quality laboratory services for robust preventive and curative care.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-218 Ensuring Deionized Water Availability for Clinical Analyzers: A Clinical Engineering Strategy for Operational Continuity in a Laboratory NetworkClinical Chemistry · 2026
- Healthcare Waste Generation Rates in the Three Categories of Public Healthcare Facilities in MalawiPublic Health Challenges · 2026
- Result harmonization in medical laboratories: accomplishments and challengesClinical Chemistry and Laboratory Medicine (CCLM) · 2026
- B-303 Evaluating an Automated Urinalysis System: Analytical Performance and Verification StrategiesClinical Chemistry · 2026
- Assessing intra- and inter-analyser imprecision in automated hematological laboratories.Journal of medical biochemistry · 2026
- Assessing the effectiveness of the integrated sample referral system in Tanzania: insights from a nationwide study.European journal of public health · 2026