Assessing intra- and inter-analyser imprecision in automated hematological laboratories.
Giuseppe Lippi, Marco Tosi, Mariateresa Rizza, Gian Luca Salvagno, Laura Pighi
University of Verona, Section of Clinical Biochemistry, Verona, Italy.
Key points
From the paper's conclusion- Sysmex XN-10 analysers exhibit strong intraand inter-analyser precision for most routine haematology tests, supporting their routine use in automated haematological lines.
Abstract
BACKGROUND: Reliable haematology results are crucial for patient diagnosis and monitoring. Maintaining low variability is particularly important for key parameters such as haemoglobin, white blood cells, and platelets, especially in automated laboratory workflows where multiple haematological analysers are connected in the same line. METHODS: Two residual whole blood samples, one normal and one pathological, were analysed in ten consecutive replicates on a Sysmex XN-10 analyser (XN-1) and then on a second connected analyser (XN-2). Intra-analyser and inter-analyser imprecision were calculated as coefficients of variation (CV%). RESULTS: Intra-analyser CVs for the normal sample ranged from 0.3% for haemoglobin to 1.5% white blood cells, while inter-analyser CVs remained below 2%. For the pathological sample, intra-analyser CVs ranged from 0.3% to 1.0%, and inter-analyser CVs reached 2.1% for hematocrit and platelets. Red blood cell count, mean corpuscular volume, and neutrophils showed CVs <1.9%. Higher variability was observed for low-abundance populations such as eosinophils and basophils (up to 20%). CONCLUSIONS: Sysmex XN-10 analysers exhibit strong intraand inter-analyser precision for most routine haematology tests, supporting their routine use in automated haematological lines.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of the HemoScreen Point-Of-Care Haematology Analyser With Sysmex XN, Beckman Coulter DxH 560 and HORIBA Yumizen H550 in Patients With Abnormal Blood Counts.International journal of laboratory hematology · 2026
- A-226 Benchtop Sysmex XN-450: Evaluation for Use in Primary Healthcare SettingClinical Chemistry · 2026
- White Blood Cell Counting in Severely Leukopenic Samples on a Sysmex XN Hematology Analyzer: Analytical Performance Evaluation of Whole Blood Mode and Low White Blood Cell Mode.Annals of laboratory medicine · 2026
- A-240 Multicenter Performance Evaluation and Method Comparison of the Yumizen H550 Hematology AnalyzerClinical Chemistry · 2026
- SỰ TƯƠNG ĐỒNG KẾT QUẢ MỘT SỐ CHỈ SỐ TẾ BÀO MÁU NGOẠI VI GIỮA MÁY ĐẾM HUYẾT HỌC TỰ ĐỘNG DXH 900 VÀ XN 10 TẠI VIỆN HUYẾT HỌC - TRUYỀN MÁU TRUNG ƯƠNGTạp chí Y học Cộng đồng · 2026
- Analytical validation of the Mindray BC-75R Vet hematology analyzer and comparison with the Siemens Advia 2120i for application in dogs and cats.Journal of veterinary diagnostic investigation : official publication of the American Association of Veterinary Laboratory Diagnosticians, Inc · 2026